Commissioning
Liquipoint FTW33
24
Hauser
10
Commissioning
10.1 Function check
Prior to commissioning, make sure that the post-installation and post-connection checks
have been performed.
See:
• "Post-installation check" checklist
• "Post-connection check" checklist
10.2 Commissioning the local display
10.2.1 Light signals (LEDs)
Position of LEDs in housing cover
4
3
1
2
2
1
3
4
A0038425
Position LED color
Description of function
1
green (gn)
Status/Communication
• lit: SIO mode
• flashing: Active communication, flash frequency
• flashes with increased luminosity: Device search (device identification), flash fre-
quency
2
yellow (ye)2
Switch status/switch output 2
lit: If the sensor is covered by medium
3
red (rd)
Warning/Maintenance required
flashing: Error remediable, e. g. invalid calibration
Fault/device failure
lit: see Diagnostics and troubleshooting
4
yellow (ye)1
Switch status/switch output 1
lit: If the sensor is covered by medium
There is no external signaling via LEDs on the metal housing cover (IP69). A
connecting cable with an M12 plug and LED indicator can be ordered as an accessory
if necessary. See "Accessories".
10.2.2 Function of LEDs
Any configuration of the switch outputs is possible. The following table shows the
behavior of the LEDs in the SIO mode: