Oxymax COS61
Wiring
Hauser
15
4
Wiring
WARNING
!
Device is energized
Improper connection can cause injury or death.
►
The electrical connection must only be carried out by a certified electrician.
►
Technical personnel must have read and understood the instructions in this manual and must
adhere to them.
►
Prior to beginning
any wiring work, make sure voltage is not applied to any of the cables.
4.1
Direct connection to the transmitter
4.1.1
Field installation
Connect the sensor directly to the transmitter by using the special measuring cable with SXP plug.
4.1.2
Panel mounting
• Remove the SXP connector (transmitter side!) from the cable.
• Refer to the following table for the cable assignment and the assigned terminals for
Liquisys COM223-WX/WS.
• Please note that the cable assignment varies depending on the sensor version (fixed cable or
TOP68 connection).
a0004118
Fig. 13: SXP
plug
Terminal
COM223
Sensor with fixed cable (OMK)
Sensor with TOP68 connection
(CYK71)
Core
Assignment
Core
Assignment
87
YE
+U
B
YE
+U
B
0
GY
0 V
WH
0 V
96
PK
Com. (digital)
GN
Communication (digital)
97
BU
Com. (digital)
BN
Communication (digital)
88
BN
–U
B
Koax innen –U
B