Installation
Oxymax COS61
8
Hauser
3.2.2
Angle of installation
The sensor can be installed up to the horizontal in an assembly, support or a suitable process
connection.
Other angles and overhead installation are not recommended. Reason: possible sediment formation
and resulting falsification of measured value.
The optimum installation angle is 45
˚
(e.g. with assembly CYA112).
a0004087
Fig. 3:
Angle of installation
A
Permissible installation positions: 0 ... 180
˚
, optimum angle 45
˚
►
Make sure you comply with the instructions for installing sensors. You will find them in the
Operating Instructions for the assembly used.
3.2.3
Mounting location
• Select the installation location so that there is easy access for later calibration.
• Make sure that upright posts and assemblies are secured safely and vibration-free.
• Select an installation location which produces a typical oxygen concentration.
3.3
Installation instructions
3.3.1
Measuring system
A complete measuring system comprises:
• Oxygen sensor Oxymax COS61
• Transmitter, e.g. Liquisys COM2x3-W
• Special measuring cable
• Assembly, e.g. COA250 flow assembly, CYA112 immersion assembly or COA451 retractable
assembly
Optional:
• CYH112 assembly holder for immersion operation
• VS junction box (for cable extension)
• Cleaning system