Proline Cubemass C 500
40
Hauser
FOUNDATION Fieldbus
2
1
3
4
7
8
6
6
6
6
5
6
6
5
A0028768
9
Connection example for FOUNDATION Fieldbus
1
Control system (e.g. PLC)
2
Power Conditioner (FOUNDATION Fieldbus)
3
Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable
specifications
4
T-box
5
Measuring device
6
Local grounding
7
Bus terminator
8
Potential matching line
Modbus RS485
2
1
A
B
3
4
4
A
B
A
B
A0028765
10
Connection example for Modbus RS485, non-hazardous area and Zone 2; Class I, Division 2
1
Control system (e.g. PLC)
2
Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable
specifications
3
Distribution box
4
Transmitter