Proline Cubemass C 500
Hauser
47
Cross-section
Cable length [max.]
0.34 mm
2
(AWG 22)
80 m (270 ft)
0.50 mm
2
(AWG 20)
120 m (400 ft)
0.75 mm
2
(AWG 18)
180 m (600 ft)
1.00 mm
2
(AWG 17)
240 m (800 ft)
1.50 mm
2
(AWG 15)
300 m (1
000 ft)
Optionally available connecting cable
Design
2 × 2 × 0.34 mm
2
(AWG 22) PVC cable
with common shield (2 pairs,
uninsulated stranded CU wires; pair-stranded)
Flame resistance
According to DIN EN 60332-1-2
Oil-resistance
According to DIN EN 60811-2-1
Shielding
Tin-plated copper-braid, optical cover ≥ 85 %
Operating temperature
When mounted in a fixed position: –50 to +105 °C (–58 to +221 °F); when cable
can move freely: –25 to +105 °C (–13 to +221 °F)
Available cable length
Fixed: 20 m (65 ft); variable: up to maximum 50 m (165 ft)
1)
UV radiation can impair the cable outer sheath. Protect the cable from direct sunshine where possible.
B: Connecting cable between sensor and transmitter: Proline 500 - digital
Standard cable
A standard cable with the following specifications can be used as the connecting cable.
Design
4, 6, 8 cores (2, 3, 4 pairs); uninsulated stranded CU wires; pair-stranded with
common shield
Shielding
Tin-plated copper-braid, optical cover ≥ 85 %
Capacitance C
Maximum 760 nF IIC, maximum 4.2 µF IIB
Inductance L
Maximum 26 µH IIC, maximum 104 µH IIB
Inductance/resistance
ratio (L/R)
Maximum 8.9 µH/Ω IIC, maximum 35.6 µH/Ω IIB (e.g. in accordance with IEC
60079-25)
Loop resistance
Power supply line (+, –): maximum 5 Ω
Cable length
Maximum 150 m (500 ft), see the following table.
Заказывайте
на
сайте
: www.tinnova.ru/endress_hauser/ ||
Эл
.
почта
: info@tinnova.ru