Proline Cubemass C 500
Hauser
49
C: Connecting cable between sensor and transmitter: Proline 500
Standard cable
6 × 0.38 mm
2
PVC cable
with common shield and individually shielded
cores
Conductor resistance
≤50 Ω/km (0.015 Ω/ft)
Capacitance: core/shield
≤420 pF/m (128 pF/ft)
Cable length (max.)
20 m (65 ft)
Cable lengths (available for
order)
5 m (15 ft), 10 m (32 ft), 20 m (65 ft)
Operating temperature
max. 105 °C (221 °F)
1)
UV radiation can impair the cable outer sheath. Protect the cable from direct sunshine where possible.
Performance characteristics
Reference operating
conditions
• Error limits based on ISO 11631
• Water with +15 to +45 °C (+59 to +113 °F) at2 to 6 bar (29 to 87 psi)
• Specifications as per calibration protocol
• Accuracy based on accredited calibration rigs that are traced to ISO 17025.
To obtain measured errors, use the
Applicator
Maximum measured error
o.r. = of reading; 1 g/cm³ = 1 kg/l; T = medium temperature
Base accuracy
Mass flow and volume flow (liquids)
±0.10 % o.r.
Mass flow (gases)
±0.50 % o.r.
Temperature
±0.5 °C ± 0.005 · T °C (±0.9 °F ± 0.003 · (T – 32) °F)
Zero point stability
DN
Zero point stability
[mm]
[in]
[kg/h]
[lb/min]
1
¹⁄₂₄
0.0008
0.00003
2
¹⁄₁₂
0.002
0.00007
4
¹⁄₈
0.014
0.0005
6
¹⁄₄
0.02
0.0007
Flow values
Flow values as turndown parameter depending on nominal diameter.