Connection data: Signal 

circuits (intrinsically safe 

circuits)

The following table contains information that depends on the transmitter type and the 

assignment of the inputs and outputs. Compare the following information to the 

information on the nameplate of the transmitter.

Terminal assignment 

Intrinsically safe values of signal circuits

Output

Basic specification, Position 3

Terminal numbers

1 (+)

2 (-)

3 (+)

4 (-)

Permanent-assignment I/O electronic modules

A

4-20 mA HART

_

B

4-20 mA HART

Impulse/Frequency/Switch output

C

4-20 mA HART

4-20 mA

G

PROFIBUS PA

Impulse/Frequency/Switch output

Output
Basic specification, Position 3

Signal circuits

Instrinsically save 
values

Electric circuits

A

4-20 mA HART

U

i

 = DC 30 V

I

i

 = 300 mA

P

i

 = 1 W

L

i

 = 0 

μ

H

C

i

 = 5 nF

Ex ia

B

4-20 mA HART

U

i

 = DC 30 V

I

i

 = 300 mA

P

i

 = 1 W

L

i

 = 0 

μ

H

C

i

 = 5 nF

Ex ia

Impulse/Frequency/

Switch output

U

i

 = 30 V

I

i

 = 300 mA

P

i

 = 1 W

L

i

 = 0 

μ

H

C

i

 = 6 nF

C

4-20 mA HART

U

i

 = DC 30 V

I

i

 = 300 mA

P

i

 = 1 W

L

i

 = 0 

μ

H

C

i

 = 30 nF

Ex ia

4-20 mA 

U

i

 = DC 30 V

I

i

 = 300 mA

P

i

 = 1 W

L

i

 = 0 

μ

H

C

i

 = 30 nF

G

PROFIBUS  PA

U

i

 = 17.5 V

I

i

 = 550 mA

P

i

 = 5.5 W

L

i

 = 10 

μ

H

C

i

 = 5 nF

or
U

i

 = 30 V

I

i

 = 300 mA

P

i

 = 1.2 W

L

i

 = 10 

μ

H

C

i

 = 5 nF

Ex ia

Impulse/Frequency/
Switch output

U

i

 = 30 V

I

i

 = 300 mA

P

i

 = 1 W

L

i

 = 0 

μ

H

C

i

 = 6 nF

Summary of Contents for Proline Promass 200

Page 1: ...ine Promass 200 NEPSI Zone 1 Zone 0 1 Zone 21 Ex i version en Document XA00156D Safety instructions for electrical apparatus for explosion hazardous areas classified according to National Supervision...

Page 2: ...Products Solutions Services...

Page 3: ...certificates of conformity 4 Extended order code 4 Safety instructions General 6 Safety instructions Installation 6 Safety instructions Zone 0 7 Safety instructions Zone 21 7 Temperature tables Compa...

Page 4: ...ty GB3836 1 4 2010 GB12476 1 2000 Certification numbers GYJ11 1524 Referential certification numbers GYJ11 1519U GYJ11 1520U GYJ11 1521U GYJ11 1522U GYJ11 1523U Inspection body NEPSI National Supervis...

Page 5: ...ial wetted parts inspection certificate More detailed information about the device is provided in the following tables These tables describe the individual positions and IDs in the extended order code...

Page 6: ...must be carried out by staff authorized to perform such work by Endress Hauser Observe all the technical data of the device see nameplate When using in hybrid mixtures gas and dust occurring simultane...

Page 7: ...ntegrate the sensor into the potential equalization system via the pipe Safety instructions Zone 0 In the event of potentially explosive vapour air mixtures only operate the device under atmospheric c...

Page 8: ...on with overvoltage protection OVP in connection with temperature class T5 T6 Ta Ta 2 C 2 For specified sensors with a maximal medium temperature Tm 200 C Tm 170 C 3 For specified sensors with a maxim...

Page 9: ...h a maximal medium temperature Tm 200 C Tm 200 C Output basic specification position 3 G Approval Basic specification positions 1 2 NEPSI NF N4 Sensor type Ta C T6 85 C T5 100 C T4 135 C T3 200 C T2 3...

Page 10: ...he maximum surface temperature Procedure for determining the maximum surface temperature 1 Select the device Promass F 200 nominal diameter DN 25 and ambient temperature Ta 60 C in the associated temp...

Page 11: ...Impulse Frequency Switch output C 4 20 mA HART 4 20 mA G PROFIBUS PA Impulse Frequency Switch output Output Basic specification Position 3 Signal circuits Instrinsically save values Electric circuits...

Page 12: ...www addresses endress com...

Reviews: