Proline Promass 200
11
Connection data: Signal
circuits (intrinsically safe
circuits)
The following table contains information that depends on the transmitter type and the
assignment of the inputs and outputs. Compare the following information to the
information on the nameplate of the transmitter.
Terminal assignment
Intrinsically safe values of signal circuits
Output
Basic specification, Position 3
Terminal numbers
1 (+)
2 (-)
3 (+)
4 (-)
Permanent-assignment I/O electronic modules
A
4-20 mA HART
_
B
4-20 mA HART
Impulse/Frequency/Switch output
C
4-20 mA HART
4-20 mA
G
PROFIBUS PA
Impulse/Frequency/Switch output
Output
Basic specification, Position 3
Signal circuits
Instrinsically save
values
Electric circuits
A
4-20 mA HART
U
i
= DC 30 V
I
i
= 300 mA
P
i
= 1 W
L
i
= 0
μ
H
C
i
= 5 nF
Ex ia
B
4-20 mA HART
U
i
= DC 30 V
I
i
= 300 mA
P
i
= 1 W
L
i
= 0
μ
H
C
i
= 5 nF
Ex ia
Impulse/Frequency/
Switch output
U
i
= 30 V
I
i
= 300 mA
P
i
= 1 W
L
i
= 0
μ
H
C
i
= 6 nF
C
4-20 mA HART
U
i
= DC 30 V
I
i
= 300 mA
P
i
= 1 W
L
i
= 0
μ
H
C
i
= 30 nF
Ex ia
4-20 mA
U
i
= DC 30 V
I
i
= 300 mA
P
i
= 1 W
L
i
= 0
μ
H
C
i
= 30 nF
G
PROFIBUS PA
U
i
= 17.5 V
I
i
= 550 mA
P
i
= 5.5 W
L
i
= 10
μ
H
C
i
= 5 nF
or
U
i
= 30 V
I
i
= 300 mA
P
i
= 1.2 W
L
i
= 10
μ
H
C
i
= 5 nF
Ex ia
Impulse/Frequency/
Switch output
U
i
= 30 V
I
i
= 300 mA
P
i
= 1 W
L
i
= 0
μ
H
C
i
= 6 nF
Summary of Contents for Proline Promass 200
Page 2: ...Products Solutions Services...
Page 12: ...www addresses endress com...