10 Technical data
Proline Promass 83 FOUNDATION Fieldbus
102
Hauser
Pressure loss coefficient for Promass F
a0001396
Fig. 39:
Pressure loss diagram for water
DN
d [m]
K
K1
K2
8
5.35
⋅
10
–3
5.70
⋅
10
7
9.60
⋅
10
7
1.90
⋅
10
7
15
8.30
⋅
10
–3
5.80
⋅
10
6
1.90
⋅
10
7
10.60
⋅
10
5
25
12.00
⋅
10
–3
1.90
⋅
10
6
6.40
⋅
10
6
4.50
⋅
10
5
40
17.60
⋅
10
–3
3.50
⋅
10
5
1.30
⋅
10
6
1.30
⋅
10
5
50
26.00
⋅
10
–3
7.00
⋅
10
4
5.00
⋅
10
5
1.40
⋅
10
4
80
40.50
⋅
10
–3
1.10
⋅
10
4
7.71
⋅
10
4
1.42
⋅
10
4
100
51.20
⋅
10
–3
3.54
⋅
10
3
3.54
⋅
10
4
5.40
⋅
10
3
150
68.90
⋅
10
–3
1.36
⋅
10
3
2.04
⋅
10
4
6.46
⋅
10
2
DN 50
DN 40
DN 25
DN 8
10000
1000
100
10
1
0.1
0.001
0.01
0.1
1
10
100
1000
DN 15
DN 80
DN 100
[mbar]
[t/h]
DN 150
DN 250
Summary of Contents for Proline Promass 83
Page 24: ...3 Installation Proline Promass 83 FOUNDATION Fieldbus 24 Endress Hauser ...
Page 34: ...4 Wiring Proline Promass 83 FOUNDATION Fieldbus 34 Endress Hauser ...
Page 64: ...6 Commissioning Proline Promass 83 FOUNDATION Fieldbus 64 Endress Hauser ...
Page 66: ...7 Maintenance Proline Promass 83 FOUNDATION Fieldbus 66 Endress Hauser ...
Page 90: ...9 Trouble shooting Proline Promass 83 FOUNDATION Fieldbus 90 Endress Hauser ...
Page 116: ...Index Proline Promass 83 FOUNDATION Fieldbus 116 Endress Hauser ...
Page 118: ...www endress com worldwide BA065D 06 en 12 04 50098623 FM SGML 6 0 ...