Proline Promass 83 FOUNDATION Fieldbus
3 Installation
Hauser
19
3.2.5
Thermal insulation
Some fluids require suitable measures to avoid loss of heat at the sensor. A wide range of materials
can be used to provide the required thermal insulation.
F06-83FHTxxx-06-00-xx-xx-001
Fig. 9:
In the case of the Promass F high-temperature version, a maximum insulation thickness of 60 mm must be
observed in the area of the electronics/neck.
If the Promass F high-temperature version is installed horizontally (with transmitter head pointing
upwards), an insulation thickness of min. 10 mm is recommended to reduce convection. The maxi-
mum insulation thickness of 60 mm must be observed (see Fig. 9).
3.2.6
Inlet and outlet runs
There are no installation requirements regarding inlet and outlet runs. If possible, install the sensor
well clear of fittings such as valves, T-pieces, elbows, etc.
3.2.7
Vibrations
The high oscillation frequency of the measuring tubes ensures that the correct operation of the
measuring system is not influenced by pipe vibrations. Consequently, the sensors require no special
measures for attachment.
3.2.8
Limiting flow
See the information on Page 91 and 100.
max. 60
Esc
E
-
+
max. 60
Summary of Contents for Proline Promass 83
Page 24: ...3 Installation Proline Promass 83 FOUNDATION Fieldbus 24 Endress Hauser ...
Page 34: ...4 Wiring Proline Promass 83 FOUNDATION Fieldbus 34 Endress Hauser ...
Page 64: ...6 Commissioning Proline Promass 83 FOUNDATION Fieldbus 64 Endress Hauser ...
Page 66: ...7 Maintenance Proline Promass 83 FOUNDATION Fieldbus 66 Endress Hauser ...
Page 90: ...9 Trouble shooting Proline Promass 83 FOUNDATION Fieldbus 90 Endress Hauser ...
Page 116: ...Index Proline Promass 83 FOUNDATION Fieldbus 116 Endress Hauser ...
Page 118: ...www endress com worldwide BA065D 06 en 12 04 50098623 FM SGML 6 0 ...