3 Installation
Proline Promass 83 FOUNDATION Fieldbus
22
Hauser
Installation in control panel
1.
Prepare the opening in the panel (Fig. 13).
2.
Slide the housing into the opening in the panel from the front.
3.
Screw the fasteners onto the wall-mount housing.
4.
Place the threaded rods in the fasteners and screw them down until the housing is seated
tightly against the panel. Afterwards, tighten the locking nuts. Additional support is not neces-
sary.
a0001131
Fig. 13:
Panel Installation (wall-mount housing)
Pipe mounting
The assembly should be performed by following the instructions in Fig. 14.
"
Caution!
If the device is mounted to a warm pipe, make certain that the housing temperature does not exceed
+60 °C, which is the maximum permissible temperature.
a0001132
Fig. 14:
Pipe mounting (wall-mount housing)
245
~1
10
+0.5 –
0.5
210
+0.5
– 0
.5
Ø
20...70
~155