Proline Promass 83 FOUNDATION Fieldbus
9 Trouble-shooting
Hauser
73
12
Device status message (FF):
Amplifier EEPROM data incon-
sistent – Err. No. 012
Local display:
S
AMP SW EEPROM
$
# 012
[.....] Quality = BAD
OUT. QUALITY = BAD
Cause:
Error when accessing data of the
measuring amplifier EEPROM
Remedy:
Perform a “warm restart” (restart the
measuring system without disconnec-
ting mains power).
• FF: Transducer Block
→
Parameter
“System - Reset”
→
RESTART
SYSTEM
• Local display:
SUPERVISION
→
SYSTEM
→
OPERATION
→
SYSTEM RESET
(
→
RESTART SYSTEM)
[.....] Substatus = Device failure
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Device needs
maintenance now
BLOCK_ERR = Input Failure
(faulty input value from Transducer
Block)
Transducer_Error = Data integrity
error
31
Device status message (FF):
S-DAT failure / not inserted –
Err. No. 031
Local display:
S
SENSOR HW DAT
$
# 031
[.....] Quality = BAD
OUT. QUALITY = BAD
Cause:
1.
S-DAT is defective.
2.
S-DAT is not plugged into amplifier
board (missing)
Remedy:
1.
Replace the S-DAT.
Spare parts
→
Check the spare part set number to
ensure that the new, replacement
DAT is compatible with the measu-
ring electronics.
2.
Plug S-DAT into amplifier board
[.....] Substatus = Device failure
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Device needs
maintenance now
BLOCK_ERR = Input Failure
(faulty input value from Transducer
Block)
Transducer_Error = Electronics
failure
32
Device status message (FF):
S-DAT data inconsistent –
Err. No. 032
Local display:
S
SENSOR SW DAT
$
# 032
[.....] Quality = BAD
OUT. QUALITY = BAD
Cause:
Error accessing the calibration values
stored in the
S-DAT.
Remedy:
1.
Check whether the
S-DAT is correctly plugged into the
amplifier board
2.
Replace the S-DAT if it is defective.
Spare parts
→
Check the spare part set number to
ensure that the new, replacement
DAT is compatible with the measu-
ring electronics.
3.
Replace measuring electronics
boards if necessary. Spare parts
[.....] Substatus = Device failure
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Device needs
maintenance now
BLOCK_ERR = Input Failure
(faulty input value from Transducer
Block)
Transducer_Error = Data integrity
error
41
Device status message (FF):
T-DAT failure – Err. No. 041
Local display:
S
TRANSM. HW-DAT
$
# 041
[.....] Quality = BAD
OUT. QUALITY = BAD
Cause:
1.
T-DAT is faulty
2.
T-DAT is not plugged into amplifier
board.
Remedy:
1.
Replace the T-DAT.
Spare parts
→
Check the spare part set number to
ensure that the new, replacement
DAT is compatible with the measu-
ring electronics.
2.
Plug T-DAT into amplifier board
→
[.....] Substatus = Device failure
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Device needs
maintenance now
BLOCK_ERR = Input Failure
(faulty input value from Transducer
Block)
Transducer_Error = Electronics
failure
No.
Device status message FF
(Diag. - Act. Syst. Cond.)
Message on local display
Transducer block
Error messages
[.....] = process variable (e.g. mass
flow, volume flow, ...)
Analog Input Funct. Block
Error messages
Cause/remedy
Summary of Contents for Proline Promass 83
Page 24: ...3 Installation Proline Promass 83 FOUNDATION Fieldbus 24 Endress Hauser ...
Page 34: ...4 Wiring Proline Promass 83 FOUNDATION Fieldbus 34 Endress Hauser ...
Page 64: ...6 Commissioning Proline Promass 83 FOUNDATION Fieldbus 64 Endress Hauser ...
Page 66: ...7 Maintenance Proline Promass 83 FOUNDATION Fieldbus 66 Endress Hauser ...
Page 90: ...9 Trouble shooting Proline Promass 83 FOUNDATION Fieldbus 90 Endress Hauser ...
Page 116: ...Index Proline Promass 83 FOUNDATION Fieldbus 116 Endress Hauser ...
Page 118: ...www endress com worldwide BA065D 06 en 12 04 50098623 FM SGML 6 0 ...