9 Trouble-shooting
Proline Promass 83 FOUNDATION Fieldbus
74
Hauser
42
Device status message
(FF):
T-DAT data inconsistent –
Err. No. 042
Local display:
S
TRANSM. SW-DAT
$
# 042
[.....] Quality = BAD
OUT. QUALITY = BAD
Cause:
Error accessing the calibration values
stored in the
T-DAT.
Remedy:
1.
Check whether the
T-DAT is correctly plugged into the
amplifier board
2.
Replace T-DAT if faulty.
Spare parts
→
Check the spare part set number to
ensure that the new, replacement
DAT is compatible with the measu-
ring electronics.
3.
Replace measuring electronics
boards if necessary. Spare parts
→
[.....] Substatus = Device failure
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Device needs
maintenance now
BLOCK_ERR = Input Failure
(faulty input value from Transducer
Block)
Transducer_Error = Data integrity
error
51
Device status message (FF):
Compatibility AMP-I/O Modul
– Err. No. 051
Local display:
S
A / C COMPATIB.
$
# 051
[.....] Quality = BAD
OUT. QUALITY = BAD
Cause:
The I/O board and the amplifier board
are not compatible.
Remedy
Use only compatible modules and
boards.
Check the compatibility of the modules
used. Check the:
– Spare part set number
– Hardware revision code
[.....] Substatus = Device failure
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Device needs
maintenance now
BLOCK_ERR = Input Failure
(faulty input value from Transducer
Block)
Transducer_Error = I/O failure
(input/output error)
111
Device status message (FF):
Totalizer checksum error –
Err. No. 111
Local display:
S
CHECKSUM TOTAL.
$
# 111
[.....] Quality = BAD
OUT. QUALITY = BAD
Cause:
Totalizer checksum error
Remedy:
1.
Restart the measuring device
2.
Replace the amplifier board if
necessary. Spare parts
[.....] Substatus = Device failure
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Device needs
maintenance now
BLOCK_ERR = Input Failure
(faulty input value from Transducer
Block)
Transducer_Error = Electronics
failure
No.
Device status message FF
(Diag. - Act. Syst. Cond.)
Message on local display
Transducer block
Error messages
[.....] = process variable (e.g. mass
flow, volume flow, ...)
Analog Input Funct. Block
Error messages
Cause/remedy
Summary of Contents for Proline Promass 83
Page 24: ...3 Installation Proline Promass 83 FOUNDATION Fieldbus 24 Endress Hauser ...
Page 34: ...4 Wiring Proline Promass 83 FOUNDATION Fieldbus 34 Endress Hauser ...
Page 64: ...6 Commissioning Proline Promass 83 FOUNDATION Fieldbus 64 Endress Hauser ...
Page 66: ...7 Maintenance Proline Promass 83 FOUNDATION Fieldbus 66 Endress Hauser ...
Page 90: ...9 Trouble shooting Proline Promass 83 FOUNDATION Fieldbus 90 Endress Hauser ...
Page 116: ...Index Proline Promass 83 FOUNDATION Fieldbus 116 Endress Hauser ...
Page 118: ...www endress com worldwide BA065D 06 en 12 04 50098623 FM SGML 6 0 ...