9 Trouble-shooting
Proline Promass 83 FOUNDATION Fieldbus
78
Hauser
388
389
390
Device status message (FF):
Amplifier error –
Err. No. 388 / 389 / 390
Local display:
S
AMP. FAULT
$
# 388 / 389 / 390
[.....] Quality = BAD
OUT. QUALITY = BAD
Cause:
Amplifier error
Remedy:
Contact your Hauser service
organisation.
[.....] Substatus = Device failure
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Device needs
maintenance now
BLOCK_ERR = Input Failure
(faulty input value from Transducer
Block)
Transducer_Error = Electronics
failure
501
Device status message (FF):
New amplifier software loaded
– Err. No. 501
Local display:
P
SW.-UPDATE ACT.
!
# 501
[.....] Quality = BAD
OUT. QUALITY = BAD
Cause:
New amplifier or communication (I/O
module) software version is loaded.
Currently no other functions are possi-
ble.
Remedy:
Wait until process is finished. The
device will restart automatically.
[.....] Substatus = Sensor failure
OUT. SUBSTATUS = Sensor
Failure
BLOCK_ERR = Device needs
maintenance now
BLOCK_ERR = Input Failure
(faulty input value from Transducer
Block)
Transducer_Error = Mechanical
failure
502
Device status message (FF):
Up-/download device data
active – Err. No. 502
Local display:
P
OSC. AMP. LIMIT
!
# 502
[.....] Quality = BAD
OUT. QUALITY = BAD
Cause:
Up- or downloading the device data via
configuration program. Currently no
other functions are possible.
Remedy:
Wait until process is finished.
[.....] Substatus = Sensor failure
OUT. SUBSTATUS = Sensor
Failure
BLOCK_ERR = Device needs
maintenance now
BLOCK_ERR = Input Failure
(faulty input value from Transducer
Block)
Transducer_Error = Mechanical
failure
586
Device status message (FF):
No fluid continuation –
Err. No. 586
Local display:
P
OSC. AMP. LIMIT
$
# 586
[.....] Quality = BAD
OUT. QUALITY = BAD
Cause:
The fluid properties do not allow a con-
tinuation of the measurement.
Causes:
– Extremely high viscosity
– Process fluid is very inhomogeneous
(gas or solid content)
Remedy:
Change or improve process conditions.
[.....] Substatus = Sensor failure
OUT. SUBSTATUS = Sensor
Failure
BLOCK_ERR = Device needs
maintenance now
BLOCK_ERR = Input Failure
(faulty input value from Transducer
Block)
Transducer_Error = Mechanical
failure
587
Device status message (FF):
Extreme process conditions –
Err. No. 587
Local display:
P
TUBE NOT OSC.
$
# 587
[.....] Quality = BAD
OUT. QUALITY = BAD
Cause:
Extreme process conditions exist. The
measuring system can therefore not be
started.
Remedy:
Change or improve process conditions.
[.....] Substatus = Sensor failure
OUT. SUBSTATUS = Sensor
Failure
BLOCK_ERR = Device needs
maintenance now
BLOCK_ERR = Input Failure
(faulty input value from Transducer
Block)
Transducer_Error = Mechanical
failu
No.
Device status message FF
(Diag. - Act. Syst. Cond.)
Message on local display
Transducer block
Error messages
[.....] = process variable (e.g. mass
flow, volume flow, ...)
Analog Input Funct. Block
Error messages
Cause/remedy
Summary of Contents for Proline Promass 83
Page 24: ...3 Installation Proline Promass 83 FOUNDATION Fieldbus 24 Endress Hauser ...
Page 34: ...4 Wiring Proline Promass 83 FOUNDATION Fieldbus 34 Endress Hauser ...
Page 64: ...6 Commissioning Proline Promass 83 FOUNDATION Fieldbus 64 Endress Hauser ...
Page 66: ...7 Maintenance Proline Promass 83 FOUNDATION Fieldbus 66 Endress Hauser ...
Page 90: ...9 Trouble shooting Proline Promass 83 FOUNDATION Fieldbus 90 Endress Hauser ...
Page 116: ...Index Proline Promass 83 FOUNDATION Fieldbus 116 Endress Hauser ...
Page 118: ...www endress com worldwide BA065D 06 en 12 04 50098623 FM SGML 6 0 ...