Replacing the device fuse
#
Warning!
Risk of electric shock. Exposed components carry dangerous voltages. Make sure that the power
supply is switched off before you remove the cover of the electronics compartment.
The main fuse is on the power unit board (Fig. 37).
The procedure for replacing the fuse is as follows:
1.
Switch off power supply.
2.
Remove the power unit board
3.
Remove cap (1) and replace the device fuse (2).
Use only fuses of the following type:
– Power supply 20...55 V AC / 16...62 V DC
→
2.0 A slow-blow / 250 V; 5.2 x 20 mm
– Power supply 85...260 V AC
→
0.8 A slow-blow / 250 V; 5.2 x 20 mm
– Ex-rated devices
→
see the Ex documentation.
4.
Assembly is the reverse of the disassembly procedure.
"
Caution!
Use only original Hauser parts.
a0001148
Fig. 37:
Replacing the device fuse on the power unit board