Proline Promass 83 FOUNDATION Fieldbus
10 Technical data
Hauser
95
Communication profiles sup-
ported
31 PS
→
Publisher, Subscriber
32 L
→
Link Master
LAS function
LAS function supported.
Galvanic isolation
All circuits for inputs, outputs, and power supply are galvanically isolated from each other.
10.1.5
Auxiliary energy
Electrical connections
Supply voltage
85...260 V AC, 45...65 Hz
20...55 V AC, 45...65 Hz
16...62 V DC
Potential equalisation
No measures necessary
Cable entries
Power supply and signal cables (inputs/outputs):
• Cable entry M20 x 1.5 (8...12 mm)
• Threads for cable entries, 1/2" NPT, G 1/2"
Connecting cable for remote version:
• Cable entry M20 x 1.5 (8...12 mm)
• Threads for cable entries, 1/2" NPT, G 1/2"
Cable specifications
remote version
• 6 x 0.38 mm
2
PVC cable with common shield and individually shielded cores.
• Conductor resistance:
≤
50
Ω
/km
• Capacitance: core/shield:
≤
420 pF/m
• Cable length: max. 20 m
• Permanent operating temperature: max. +105 °C
Power consumption
AC: <15 VA (including sensor)
DC: <15 W (including sensor)
Switch-on current:
• max. 13.5 A (< 50 ms) at 24 V DC
• max. 3 A (< 5 ms) at 260 V AC
Power supply failure
Lasting min. 1 power cycle:
• EEPROM or T-DAT saves measuring system data if power supply fails.
• S-DAT is an exchangeable data storage chip with sensor specific data: nominal diameter, serial
number, calibration factor, zero point, etc.
Summary of Contents for Proline Promass 83
Page 24: ...3 Installation Proline Promass 83 FOUNDATION Fieldbus 24 Endress Hauser ...
Page 34: ...4 Wiring Proline Promass 83 FOUNDATION Fieldbus 34 Endress Hauser ...
Page 64: ...6 Commissioning Proline Promass 83 FOUNDATION Fieldbus 64 Endress Hauser ...
Page 66: ...7 Maintenance Proline Promass 83 FOUNDATION Fieldbus 66 Endress Hauser ...
Page 90: ...9 Trouble shooting Proline Promass 83 FOUNDATION Fieldbus 90 Endress Hauser ...
Page 116: ...Index Proline Promass 83 FOUNDATION Fieldbus 116 Endress Hauser ...
Page 118: ...www endress com worldwide BA065D 06 en 12 04 50098623 FM SGML 6 0 ...