Proline Prosonic Flow P
Installation
Hauser
31
• Two sensors incl. connecting cables
• Open-ended wrench (13 mm)
• Screwdriver
Procedure:
1. Position the strapping bands using the mounting rail [only DN50 to 600 (2 to 24"), for
larger nominal diameters, measure the distance between the center of the strap bolts
directly]: Fit the mounting rail with the bore identified by the letter (from the
Result
sensor distance / measuring aid
parameter) over the mounting bolt of strapping band
1 that is fixed in place. Position the adjustable strapping band 2 and fit the mounting
rail with the bore identified by the numerical value over the mounting bolt.
G
22
A0043384
20
Determining the distance according to the mounting rail (e.g. G22)
2. Tighten strapping band 2 so that it cannot slip.
3. Remove the mounting rail from the mounting bolt.
4. Fit the sensor holders over the individual mounting bolts and tighten securely with the
locking nut.
5. Screw the mounting rail holders onto the sensor holder.
Summary of Contents for Proline Prosonic Flow P
Page 39: ......