10. Insert the sensor cable into the sensor until the end stop.
A0043386
23
Mounting the sensor and connecting the sensor cable
The sensors can now be connected to the transmitter via the sensor cables and the
error message can be checked in the sensor check function. This completes the
mounting procedure.
• The visible measuring pipe surface must be clean (free from flaking paint and/or rust)
to ensure good acoustic contact.
• If the sensor is removed from the measuring pipe, it must be cleaned and new coupling
gel applied (if there is no coupling pad).
• On rough measuring pipe surfaces, the gaps in the rough surface must be filled with
sufficient amounts of coupling gel if the use of the coupling pad does not suffice
(installation quality check).
5.2.5
Mounting the transmitter housing
L
CAUTION
Ambient temperature too high!
Danger of electronics overheating and housing deformation.
‣
Do not exceed the permitted maximum ambient temperature .
‣
If operating outdoors: Avoid direct sunlight and exposure to weathering, particularly in
warm climatic regions.
L
CAUTION
Excessive force can damage the housing!
‣
Avoid excessive mechanical stress.
The transmitter can be mounted in the following ways:
• Post mounting
• Wall mounting