Prosonic S FDU90/91/91F/92/93/95/96
14
Hauser
Note!
• The process seal is not included in the delivery.
• Hauser supplies DIN/EN flanges made of stainless steel AISI 316L with the material number
1.4404 or 1.4435. With regard to their temperature stability properties, the materials 1.4404 and 1.4435
are grouped under 13E0 in EN 1092-1 Tab. 18. The chemical composition of the two materials can be
identical.
Caution!
For 3A applications:
The internal diameter of the nozzle should be selected according to the valid allowable limits for 3A
applications. Usually, the internal diameter of the nozzle should be larger than or equal to the internal diameter
of the sensor.
Nozzle installation
Install the sensor at a height so that the blocking distance BD is not undershot, even at maximum fill level. Use
a pipe nozzle if you cannot maintain the blocking distance in any other way. The interior of the nozzle must
be smooth and may not contain any edges or welded joints. In particular, there should be no burr on the inside
of the tank side nozzle end. Note the specified limits for nozzle diameter and length. To minimise disturbing
factors, we recommend an angled socket edge (ideally 45°).
L00-FDU9xxxx-17-00-00-xx-006
Maximum nozzle length [mm (in)]
Nozzle
diameter
FDU90
1)
1)
mounted at the rear side thread
FDU90
2)
2)
mounted at the front side thread (flush mounting)
FDU91
FDU91F
FDU92
FDU93
FDU95
FDU96
DN50/2"
50 (1.97)
DN80/3"
340 (13.4) 250 (9.84) 340 (13.4) 250 (9.84)
3)
3)
Valid for flush mounting; for mounting with G/NPT 1" and DN100 or higher see FDU91.
DN100/4"
390 (15.4) 300 (11.8) 390 (15.4) 300 (11.8)
DN150/6"
400 (15.7) 300 (11.8) 400 (15.7) 300(11.8)
400 (15.7)
DN200/8"
400 (15.7) 300 (11.8) 400 (15.7) 300(11.8)
400 (15.7) 520 (20.5)
DN250/10"
400 (15.7) 300 (11.8) 400 (15.7) 300(11.8)
400 (15.7) 520 (20.5) 630 (24.8)
DN300/12"
400 (15.7) 300 (11.8) 400 (15.7) 300(11.8)
400 (15.7) 520 (20.5) 630 (24.8) 800 (31.5)
Sensor characteristics
Emission angle
12°
12°
9°
12°
11°
4°
5°
6°
Blocking
distance [m (ft)]
0.07 (0.2)
0.07 (0.2)
0.3 (1)
0.3 (1)
0.4 (1.3)
0.6 (2)
0.7 (2.3)
1.6 (5.2)
Max. measuring
range [m (ft))
in liquids
3 (9.8)
3 (9.8)
10 (33)
10 (33)
20 (66)
25 (82)
Max. measuring
range [m]
in solids
1.2 (3.9)
1.2 (3.9)
5 (16)
5 (16)
10 (33)
15 (49)
45 (148)
70 (230)
L
D
FDU9x
D
FDU91F
45°
L
L
D
L
D
FDU90
FDU90
45°
45°
45°