Prosonic S FDU90/91/91F/92/93/95/96
16
Hauser
Environment
Ingress protection
Tested according to IP68/NEMA6P (24 h at 6 ft under water surface)
Vibration resistance
DIN EN 600068-2-64; 20 to 2000 Hz; 1 (m/s
2
)
2
/Hz; 3x100 min.
Storage temperature
Identical to process temperature, see below
Thermal shock resistance
According to DIN EN 60068-2-14; examination to min/max process temperature; 0.5 K/min; 1000 h
Electromagnetic compatibility
Electromagnetic compatibility according to all relevant requirements of the EN 61326- series and NAMUR
recommendation EMC (NE21). For details see declaration of conformity.
With respect to interference emission the devices meet the requirements of class A and are only provided for
use in an "industrial environment"!
Process
Process temperature,
Process pressure
Sensor
Process temperature
Process pressure (abs.)
FDU90
40 to +80 °C (
40 to +176 °F)
0.7 to 4 bar (10.15 to 58 psi)
FDU91
40 to +80 °C (
40 to +176 °F)
1)
1)
In order to avoid ice build-up, the sensors FDU90 and FDU91 are available in a version with integrated sensor
heater (
6). If this heater is used, an external temperature sensor has to be applied for time-of-flight
correction. The transmitter FMU90 is available in a version with an input for the external temperature sensor.
For details refer to Technical Information TI00397F.
0.7 to 4 bar (10.15 to 58 psi)
FDU91F
-40 to +105 °C (-40 to +221 °F)
(30 min/135 °C (275 °F))
2)
for Ex instruments: -40 to +80 °C (-40 to +176 °F)
2)
Only valid for Tri-clamp and flush mounting
0.7 to 4 bar (10.15 to 58 psi)
FDU92
40 to +95 °C (
40 to+203 °F)
for Ex instruments: -40 to +80 °C (-40 to +176 °F)
0.7 to 4 bar (10.15 to 58 psi)
FDU93
40 to +95 °C (
40 to +203 °F)
for Ex instruments: -40 to +80 °C (-40 to +176 °F)
0.7 to 3 bar (10.15 to 43.5 psi)
FDU95 - *1***
(low temperature version)
40 to +80 °C (
40 to +176 °F)
0.7 to 1.5 bar (10.15 to 22 psi)
FDU95 - *2***
(high temperature version)
40 to +150 °C (
40 to +302 °F)
for Dust-Ex versions:
40 to +130 °C
0.7 to 1.5 bar (10.15 to 22 psi)
FDU96
40 to +150 °C (-40 to +302 °F)
for Dust-Ex or Gas-Ex versions:
40 to 140 °C (
40 to +284 °F)