Installation
6.1
Installation conditions
6.1.1
Mounting position
Mounting location
h
A0032998
7
Mounting location
Preferably install the sensor in an ascending pipe, and ensure a sufficient distance to the
next pipe elbow: h ≥ 5 x DN.
Inlet and outlet runs
Observe the following inlet and outlet runs to comply with accuracy specifications:
≥ 10 × DN
≥ 5 × DN
A0035458
8
Inlet and outlet runs
6.1.2
Environment and process requirements
Ambient temperature range
Sensor
0 to +60 °C (+32 to +140 °F)
Transmitter
0 to +60 °C (+32 to +140 °F)
If operating outdoors:
• Install the measuring device in a shady location.
• Avoid direct sunlight, particularly in warm climatic regions.
• Avoid direct exposure to weather conditions.