Teqwave H
Installation
Hauser
23
1. Unscrew the mounting plate.
2. Guide the holder rods through the pre-drilled openings in the mounting plate and
screw the mounting plate back onto the transmitter.
3. Secure the transmitter holder to a post with the clamps.
!
max. 60 (2.36)
109 (4.29)
170 (6.69)
A0044740
6.3
Post-installation check
Perform the following checks after the installation:
Device condition and specification
Note
Is the device damaged (visual inspection)?
Does the device meet the specifications, such as
• Ambient temperature
• Nominal pressure
• Measuring range → 80
Installation
Note
Has the correct mounting location been selected? → 20
Process environment/process conditions
Note
Are the specifications for the inlet run observed?
Inlet run ≥ 10 × DN
Are the specifications for the outlet run observed?
Outlet run ≥ 5 × DN
Is the measuring device protected against precipitation and direct sunlight?