Teqwave H
Electrical connection
Hauser
25
Terminal
Assignment
Description
1
Signal ground
alarm
Max. AC 30 V/DC 50 V, 1 A
Switch output
7.1.3
Requirements for the supply unit
Supply voltage
DC 24 V ±20%
Version
Touch-safe circuit according to DIN EN 61010-1, as terminal V- is electrically
connected to the transmitter housing.
Power unit
The power unit must be tested to ensure it meets safety requirements (PELV), as the
measuring device is Class III equipment.
7.2
Connecting the measuring device: transmitter with
aluminum housing
L
DANGER
Incorrect connection may result in injury or death
‣
Have electrical connection work carried out by appropriately trained specialists only.
‣
Observe applicable federal/national installation codes and regulations.
‣
Comply with local workplace safety regulations.
7.2.1
Connecting the connecting cable
Connect the sensor to the transmitter only using the connecting cable supplied.
NOTICE
Measuring device malfunction
Damaged cables can impact the functional integrity of the measuring device.
‣
Avoid strain when laying the connecting cable.
‣
Do not bend or shorten the connecting cable.
‣
Do not remove the plug of the connecting cable.
‣
Replace damaged or broken cables immediately.
‣
Route connecting cables separate from cables that carry very large electrical current
(e.g. converter and motor connecting cables).
10
Connecting the connecting cable
‣
Insert the push-pull connectors of the connecting cable into the sockets at the position
indicated until they engage with a click.
Red dots on the connectors indicate the position.