Teqwave H
Electrical connection
Hauser
27
NOTICE
Measuring device malfunction
Damaged cables can impact the functional integrity of the measuring device.
‣
Avoid strain when laying the connecting cable.
‣
Do not bend or shorten the connecting cable.
‣
Do not remove the plug of the connecting cable.
‣
Replace damaged or broken cables immediately.
‣
Route connecting cables separate from cables that carry very large electrical current
(e.g. converter and motor connecting cables).
A0043295
1. Push the connecting cable through the cable entry. To ensure tight sealing, do not
remove the sealing ring from the cable entry.
2. Insert the push-pull connectors of the connecting cable into the sockets at the
position indicated until they engage with a click. Red dots on the connectors indicate
the position.
3. Firmly tighten the cable glands.
7.3.3
Connecting the signal and supply voltage cables
The measuring device does not have an internal circuit breaker. For this reason,
assign the measuring device a switch or power-circuit breaker that allows the power
supply line to be easily disconnected from the mains.
The signal can be transmitted using analog technology via the analog output and using
digital technology via Ethernet (Modbus protocol). The connection to the "Teqwave Viewer"
operating tool is also established via the Ethernet interface. For information on
establishing the connection to the "Teqwave Viewer", see the → 34.
L
DANGER
Serious or fatal injury from electric shock
‣
Ground terminal V- and the sensor housing separately from one another because
terminal V- is connected to the sensor housing.