Teqwave H
Commissioning
Hauser
49
Parameters
Procedure
Selection/input
Factory setting
Switch point Min
Prerequisite:
Range mode
is selected in the
Output mode
parameter.
Set the lower limit at which the
relay changes its switch status,
while paying attention to the
measuring range limits → 80
and the concentration app data
sheet.
Positive decimal
• Analysis parameter 1…n:
depends on the selected
concentration app
• Temperature: 0.00
• Speed of sound: 500.00
• Density: 500.00
Hysteresis
Enter the tolerance value for the
switch points.
Entering a hysteresis prevents
the relay from switching back
and forth around the upper
and lower limits. The unit for
the hysteresis is identical to
the unit of the measured
variable.
Positive decimal
0.00
NC contact/NO contact
Select the switching behavior of the
relay.
• NC contact
• NO contact
NC contact
Output
Select the measured variable to
which the relay output should react.
• Analysis parameter 1…n
• Temperature
• Speed of sound
• Density
• Dispersion
• Deactivate relay output (can only
be selected via the transmitter)
Temperature
10.4.7 Configuring the measured value display
The measured value display is configured via various menus in the transmitter with a
touch screen and in the Viewer.
Navigation using transmitter with touch screen
1. Settings menu → "View settings"
2. Settings menu → "View settings" → "View"
3. Settings menu → "Application parameters" → "Averaging" → Select measured variable
4. Settings menu → "Application parameters" → "Averaging" → Select the "Temperature"
measured variable → "Kalman filter"
Navigation using the Viewer
1. Menu "Teqwave Transmitter" → "View settings''
2. Menu "Teqwave Transmitter" → "Averaging" → Select measured variable
3. Menu "Teqwave Transmitter" → "Averaging" → Select the "Temperature" measured
variable → "Kalman filter"
The operating tool communicates the settings to the transmitter as soon as the
Apply
button is clicked.