Technical data
Teqwave H
84
Hauser
16.5 Power supply
Terminal assignment
Supply voltage
Transmitter
DC 24 V ±20%
The power unit must be tested to ensure it meets safety requirements (PELV).
Power consumption
Transmitter
4 W
Current consumption
Transmitter
Maximum switch-on current
6 A
Power supply failure
The configuration and recorded data are retained in the device memory.
Electrical connection
Potential equalization
Terminals
Terminal type
Screw terminals
Conductor cross-
section
0.129 to 1.31 mm
2
(16 to 26 AWG)
Cable specification
16.6 Performance characteristics
Maximum measured error
Speed of sound
±2 m/s (±6.56 ft/s)
Temperature
±0.5 K
Acoustic density
±0.01 g/cm³
Repeatability
Speed of sound
±0.3 m/s (0.98 ft/s)
Temperature
±0.1 K
Acoustic density
±0.0015 g/cm³
Accuracy
Concentration measurement accuracy
The measuring device can achieve a concentration accuracy of up to 0.01 %. The accuracy
depends on the concentration app. The data sheet for this and the Applicator contain
detailed accuracy information.
Compensation of cross-interference
Depending on the concentration app, various compensation values can be entered or
read into the measuring device to compensate for known cross-interference. → 54.