Commissioning (via operating menu)
Levelflex FMP53 HART
42
Hauser
8.
Setup → Level
Displays the measured level L.
9.
Setup → Distance
Displays the distance D between the reference point R and the level L.
10.
Setup → Signal quality
Displays the signal quality of the level echo.
11.
Setup → Mapping → Confirm distance
Compare the displayed distance to the real distance in order to start the recording
of the mapping curve.
9.6
User-specific applications
For details of setting the parameters of user-specific applications, see separate
documentation:
BA01002F (Operating Instructions FMP53, HART)
For the
Expert
submenu refer to:
GP01000F (Description of Device Parameters FMP5x, HART)
Summary of Contents for HART Levelflex FMP53