INSTANT FLEXIBLE DIGITAL
THERMOMETER
USER’S
MANUAL
*Read instructions carefully before using the Digital
Thermometer.
WARNING
Choking Hazard:
Thermometer cap may be fatal when swallowed. Do not
allow children to use this device without parental
supervision.
Do not use thermometer in ear. Designed for use in oral,
rectal and armpit (axilla) readings only.
Do not place thermometer near extreme heat, it may
explode.
Use of the Probe Cover may result in a 0.2
o
F (0.1
o
C)
discrepancy from actual temperature.
Remove battery from the device when not in operation
for a long time.
The use of temperature readings for self-diagnosis is
dangerous. Consult your doctor for the interpretations of
the results. Self-diagnosis may lead to the worsening of
the existing disease conditions.
Do not attempt to get measurements when the
thermometer is wet as inaccurate readings may result. Do
not bite the thermometer. Doing so may lead to breakage
and/or injury.
Do not attempt to disassemble or repair the
thermometer. Doing so may result in inaccurate
readings.
After each use, disinfect the thermometer especially in
case the device is used by more than one person.
Do not force the thermometer into the rectum. Stop
insertion and abort the measurement when pain is
present. Failure to do so may lead to injury.
Do not use thermometer orally after being used rectally.
We advise you to purchase our Probe Covers for better
hygiene when used by the whole family.
For children who are two years old or younger, please do
not use the devices orally if the unit has been stored at
temperatures over 41
℉
~
104
℉
(5
℃
~
40
℃
), leave it in
41
℉
~
104
℉
(5
℃
~
40
℃
) ambient temperature for about
15 minutes before using it.
This digital thermometer provides a quick and highly accurate reading of an individual's body
temperature. The digital thermometer is intended to measure the human body’s temperature in
regular mode orally, rectally or under the arm, and the device is reusable for clinical or home use
on people of all ages. To better understand its functions and to provide years of dependable results,
please read all instructions first.
This appliance conforms to the following standards:
ASTM E1112 Standard Specification for Electronic Thermometer for Intermittent Determination of
Patient Temperature, ISO 80601-2-56 Medical electrical equipment —Part 2-56:Particular
requirements for basic safety and essential performance of clinical thermometers for body
temperature measurement, IEC 60601-1-11 Medical electrical equipment —Part 1-11: General
requirements for basic safety and essential performance –Collateral Standard: Requirements for
medical electrical equipment and medical electrical systems used in the home healthcare
environment and complies with the requirements of IEC 60601-1-2(EMC), AAMI/ANSI ES60601-
1(Safety) standards. And the manufacturer is ISO 13485 certified.
DIRECTIONS
1.
Press the On/Off Button next to LCD display. A tone will sound as the screen shows,
followed by last recorded temperature. After showing the self-test temperature, the
thermometer is now in the testing mode.
2.
Position the thermometer in desired location (mouth, rectum, or armpit).
a.
Oral Use
: Place thermometer under tongue as indicated by “
” position shown in
Figure 2
. Close your mouth and breathe evenly through the nose
to prevent the measurement from being influenced by
inhaled/exhaled air. Normal temperature between 96.3
o
F and
99.1
o
F (35.7
o
C and 37.3
o
C).
b. Rectal Use:
Lubricate probe tip with petroleum jelly for easy
insertion. Gently insert sensor approximately 1 cm (less than ½”)
into rectum. Normal temperature between 97.2
o
F and 99.9
o
F
(36.2
o
C and 37.7
o
C).
c. Armpit Use:
Wipe armpit dry. Place probe in armpit and keep arm pressed firmly at
side. From a medical viewpoint, this method will always provide inaccurate readings, and
should not be used if precise measurements are required. Normal temperature between
95.4
o
F and 98.1
o
F (35.2
o
C and 36.7
o
C).
3.
The degree sign flashes throughout the testing process. When flashing stops, a soft alarm
will beep. The reading may take from 10 to 30 seconds depending on where the computer
is used. The measured reading will appear on the LCD simultaneously. The minimum
measurement time until the signal tone (beep) must be maintained without exception.
The measurement continues even after the buzzing notification. So that in order to
achieve better body temperature measurement result, recommend to keep the probe in
mouth and rectum about 2 minutes, or in armpit about 5 minutes regardless of the beep
sound and at least 30 seconds measurement interval should be maintained in the first few
uses.
4.
To prolong battery life, press the On/Off Button to turn unit off after testing is complete.
If no action is taken, the unit will automatically shut off after around 10 minutes.