background image

TROUBLE SHOOTING 

 

 
HOW TO CALIBRATE YOUR THERMOMETER 

The thermometer is initially calibrated at the time of manufacturing. If the thermometer 
is  used  according  to  the  manual,  periodic  readjustment is not  required.  However,  we 
recommend conducting a quick calibration to confirm that your unit didn't get damaged 
during shipment. A quick way to calibrate your unit is by doing the water test. Dip the 
tip of the thermometer into a cup of cold water for a few seconds. Your display should 
show  'Lo'  (meaning  low  temperature,  not  low  battery)  or  a  low-temperature  reading 
number (for example it shows 33 °C or 91°F). Power it OFF then let it sit for two minutes 
or three, power it ON, dip the tip into a cup  of hot water this time for a few seconds. 
Wait for your thermometer to show 'Hi' or a high-temperature reading number on its 
LCD  display  (for  example  it  shows  38  °C  or  100  °F).  Power  it  OFF,  Leave  it  at  room 
temperature  for  30  to  45  minutes  and  start  using  it,  and  it  will  mean  that  you  have 
successfully  checked your  thermometer  and  that  you  are  good  to  go! If you  find  that 
your  reading  is  very  low  and/or  not  changing  the  display  of  the  number  on  the  LCD 
screen at all, please get in touch with us through email with your Amazon Order ID. 
 

SYMBOL EXPLANATION 

 

 
CLEANING & DISINFECTION 

Wipe  the  thermometer  with  a  soft  clean  cloth.  For  stubborn  stains,  wipe  the 
thermometer  with  a  cloth  that  has  been  dampened  in  water  or  in  neutral  detergent 
solution  and  then  wring  thoroughly.  Finish  by  wiping  with  a  soft  dry  cloth.  For 
disinfection,  75%  Ethanol  or  Isopropyl  alcohol  can  be  used.  Observe  the  following  to 
prevent damage to the thermometer: 
 

 Do not use benzene, thinner, gasoline or other strong solvents to clean the       

thermometer. 

   Do not attempt to disinfect the sensing section (tip) of the thermometer by immersing    

in alcohol or in hot water (water over 122oF/50oC). 

  Do not use ultrasonic washing to clean the thermometer. 

 

o

F/

o

C SWITCHABLE 

Temperature readings are available in the Fahrenheit or Celsius scale (

o

F/

o

C; located in 

the upper right corner of LCD). By default, the thermometer is set to 

o

F. If you want to 

switch to 

o

C, turn the unit off, press and hold On/Off Button for approximately 2 seconds 

to  change  the  current  setting.  To  go  back  to 

o

F,  power  off,  and  press  and  hold  for  2 

seconds again. It will change back to 

o

F. 

 

          

               

             

 

 

SPECIFICATIONS               

 

 

BATTERY REPLACEMENT 

1.

 

Replace battery when 

 appears in the  

lower right corner of LCD display. 
 

2.

 

Pull battery cover off as shown in Figure 3. 
 

3.

 

Gently  pull  out  plastic  circuit  board  with 
battery  chamber  approximately  1  cm 
(slightly less than 

½”

.) (See Figure 4) 

4.

 

Use  pointed  object  such  as  a  pen  to 
remove  old  battery.  Discard  battery 
lawfully. Replace with new 1.5 V DC button 
type LR41 or SR41, UCC392, or equivalent. 
Be sure battery is installed with  

“+” p

olarity facing up. (See Figure 5) 

5.

 

Slide battery chamber back into place and 
attach cover. 

 
 
 
 
 

All rights reserved.  

HAPPY CARE BY ENJI 

info@enji-prime.com

  

www.enji-prime.com

 

Enji Prime Copyright 2020 

 

 
 
 
 
 

Reviews: