ISA-40M Oxygen Monitor

ENMET

 Corporation

1

1.0 Introduction

1.0 Introduction

The 

ISA-40M

 system is designed to monitor the oxygen content of atmospheres in the vicinity of MRI (Nuclear

Magnetic Resonance) equipment.  Air contains 20.9% oxygen by volume; if oxygen in the air is displaced by a liquid
cryogen (such as nitrogen) leak in MRI equipment, the oxygen deficient atmosphere could become hazardous to
human life.  To minimize the effects of magnetic fields generated by MRI equipment, LED's and brass hardware are
used in the 

ISA-40M

.

The 

ISA-40M

 system employs galvanic micro-fuel cells as oxygen detection transducers.  The output current of each

cell increases or decreases according to the partial pressure of oxygen in the surrounding air, which corresponds to the
oxygen content.  The system monitors the oxygen content continuously, and if the oxygen content at any sensor location
drops to the alarm point (usually 18.0% by volume), the instrument provides audio and visual alarms at the control unit.
An audio alarm also occurs at the sensor location.  Internal relay contacts can be used to activate additional alarm signals
and/or external equipment.  Oxygen content from 16% to 26% by volume is constantly displayed by meters on the
control unit.  The 

ISA-40M

 system is an alarm device which minimizes the risk of harm or death to personnel by

warning them of a hazardous oxygen deficient atmosphere with enough time for them to take action.

An 

ISA-40M

 system consists of a number of identical channels of oxygen deficiency detection.  Each circuit cell

combination makes up a separate channel. 

ISA-40M-2

 has two identical circuits with related oxygen cells; likewise,

an 

ISA-40M-4

 has four channels.  If the cell is enclosed in a remote sensor, each cell can be located in a separate area

up to 1000 feet from the instrument enclosure.  The cells are connected to the control unit electrically by wire or cable.
Each plug-in circuit board contains the circuitry for two channels. 

ISA-40M

 instruments described in this manual can

contain up to 4 channels.  This particular manual is for the 

ISA-40M-1

 through the 

ISA-40M-4

.

N

OTE

:

All specifications stated in this manual may change without notice.

1.1 Unpack

Unpack the 

ISA-40M

 and examine it for shipping damage.  If such damage is observed, notify both 

ENMET

 customer service

personnel and the commercial carrier involved immediately.

Regarding Damaged Shipments

N

OTE

:

 It is your responsibility to follow these instructions.  If they are not followed, the carrier will

not honor any claims for damage.

q

 

This shipment was carefully inspected, verified and properly packaged at our company and delivered to the carrier in good condition.

q

 

When it was picked up by the carrier at 

ENMET

, it legally became your company’s property.

q

 

If your shipment arrives damaged:

 

Keep the items, packing material, and carton “As Is.”  Within 5 days of receipt, notify the carrier’s local office and request
immediate inspection of the carton and the contents.

 

After the inspection and after you have received written acknowledgment of the damage from the carrier, contact 

ENMET

 Customer

Service for return authorization and further instructions.  Have your Purchase Order and Sales Order numbers available.

q

 

ENMET

 either repairs or replaces damaged equipment and invoices the carrier to the extent of the liability coverage, usually

$100.00.  Repair or replacement charges above that value are your company’s responsibility.

q

 

The shipping company may offer optional insurance coverage. 

ENMET

 only insures shipments with the shipping company when

asked to do so in writing by our customer.  If you need your shipments insured, please forward a written request to 

ENMET

Customer Service.

Regarding Shortages

If there are any shortages or questions regarding this shipment, please notify 

ENMET

 Customer Service within 5 days of receipt at

the following address:

ENMET Corporation

680 Fairfield Court

Ann Arbor, MI 48108

734-761-1270

734-761-3220 Fax

1.2 Check Order

Check, the contents of the shipment against the purchase order.  Verify that the 

ISA-40M

 is received as ordered.  If there are

accessories on the order, verify that they are present.  Check the contents of calibration kits.  Notify 

ENMET

 customer service

personnel of any discrepancy immediately.

1.3 Serial Numbers

Each 

ISA-40M

 is serialized.  These numbers are on tags on the equipment and are on record in an 

ENMET

 database.

Summary of Contents for 04615-001

Page 1: ...Arbor MI 48106 0979 80015 102 revised 08 16 91 MCN 204 04 16 99 MCN 256 06 07 01 MCN 270 12 17 01 ISA 40M Oxygen Monitor For MRI Scanner Applications For P N s 04615 001 04615 101 04615 002 04615 102...

Page 2: ...Atmospheric Pressure Variations 10 4 4 Temperature Variations 10 5 0 MAINTENANCE 11 5 1 Cell Aging 11 5 2 Cell End of Life 11 5 3 Circuit Alignment 11 5 4 Circuit Breakers Fuse 13 6 0 ISA 40M SPECIFI...

Page 3: ...through the ISA 40M 4 NOTE All specifications stated in this manual may change without notice 1 1 Unpack Unpack the ISA 40M and examine it for shipping damage If such damage is observed notify both EN...

Page 4: ...l Alarms LED Red Visual alarm steady light When this LED is on power to the unit is on and the unit is in high alarm LED Green When this is on the unit is operating and is not in alarm When this LED i...

Page 5: ...ting the electronics control unit See Figure 2 for mounting dimensions NOTE All dimensions are given in inches 2 Apply 110 VAC and or 12 VDC power to the appropriate terminals on TB5 and TB3 Refer to...

Page 6: ...oldering iron 18 gauge or larger insulated wire and 60 tin 40 lead resin sore solder CAUTION This procedure should be performed only by personnel with electrical repair experience 1 Disconnect power f...

Page 7: ...ackage 3 Remove the metal shorting clip or spring from cell pins See Figure 3 4 Plug the cell into the component side of the cell circuit board center pin to center hole See Figure 3 Place the board i...

Page 8: ...xygen Cell TB1 2 O2 TBS 3 2 Oxygen Cell TB1 2 O2 TBS 4 3 Oxygen Cell TB2 3 O2 TBS 3 3 Oxygen Cell TB2 3 O2 TBS 4 4 Oxygen Cell TB2 4 O2 TBS 3 4 Oxygen Cell TB2 4 O2 TBS 4 Figure 4 Interior View of Sen...

Page 9: ...horting clip or spring from the cell See Figure 4 8 Plug the cell into the component side of the cell circuit board in the oxygen sensor assembly insert the center pin of the cell into the center hole...

Page 10: ...y contacts to an inappropriate load See Wiring Information Tables 3 4 and Sections 3 3 1 and 3 3 2 and the Figure 6 Utilizing Relay Contacts Internal view of ISA 40M Suggested wiring configuration for...

Page 11: ...4 TB4 4COM Common 3 3 2 Latching Relays Latching relays are available on TB7 the terminal strip located on the auxiliary relay circuit assembly located on the upper surface of the control unit enclosu...

Page 12: ...ow 32 F 0 C or above 130 F 55 C Never adjust the oxygen cell in an area where an oxygen deficiency may exist Always adjust the oxygen cell in fresh air Use bottled air if necessary Avoid exposure to t...

Page 13: ...mote sensor assembly with audio alarm disconnect the ground side of the sensor assembly audio alarm during this process to avoid spurious alarms while changing the cell reconnect it when done 5 3 Circ...

Page 14: ...e the oxygen cell is properly connected by measuring the voltage between the cell leads this voltage must be between 0 043 and 0 09 VDC c Measure the Low Level Set voltage for the channel by connectin...

Page 15: ...ttled 20 9 oxygen to the sensor for a couple of minutes then adjust gain e If possible exhale all your breath over the oxygen cell The meter reading should drop and the oxygen alarm should come on at...

Page 16: ...ating Temp Range 32 F to 125 F ENMET Warranty 6 months form shipment Size 1 25 diameter by 0 80 long Weight 26 grams Expected Life Up to 14 months in air NOTE All specifications stated in this manual...

Page 17: ...information can result in processing delays ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour All instruments sent to ENMET...

Page 18: ...e difference between the most economical means and the expedient mode ENMET shall not be liable for any loss or damage caused by the improper use of the product The purchaser indemnifies and saves har...

Reviews: