AM-5175
ENMET Corporation
9
5.0 Maintenance
The
AM-5175
maintenance menus are accessed by pressing the
M
ENU
button and
S
ELECT
button as described in the
maintenance menu section.
5.1 Maintenance Menus
C
AUTION
:
Do Not Attempt A Span Procedure Without Calibration Gas Applied to The Sensor; if this is done, the instrument is
forced into a calibration fault mode.
Pushbutton switches control the
M
ENU
and
S
ELECT
functions. The
M
ENU
and
S
ELECT
button locations are indicated on the
display panel, see Figure 1. The
M
ENU
button is used to display the various menu options and make incremental changes to
numbers such as alarm points, calibrations gas, etc. The
S
ELECT
button is used to select that option, set zero or span digit.
To enter the maintenance menu press and hold the
M
ENU
button for 2 to 4 seconds
Table 2 indicates the maintenance menu sequence see Figure 6 for a detailed maintenance menu flow chart.
N
OTE
:
Software revision may cause variations of display output.
Table 2: AM-5175 Maintenance Menus Sequence
Example of Display
Function
Normal Display Mode
Measurement of CO
Press and hold the
M
ENU
button for 2 – 4 seconds to enter the Maintenance Menu
The Power/Fault LED will flash Green – Red to indicate the
AM-5175
is in Maintenance Mode
To exit the maintenance Menu and return to the
Normal Display Mode:
If intended function Press
S
ELECT
button
Press the
M
ENU
button to advance to the Zero procedure
Not Present for Oxygen Units
For adjusting Zero:
If intended function Press
S
ELECT
button
Press the
M
ENU
button to advance to the Span procedure
For adjusting the Span:
If intended function Press
S
ELECT
button
Press the
M
ENU
button to advance to each Alarm set point procedures
For adjusting the Alarm 1, 2 and 3 set points:
If Intended function Press
S
ELECT
button
Press the
M
ENU
button to advance the mA Span set point procedure
For adjusting the mA Span set point:
If intended function Press
S
ELECT
button
Pressing the
M
ENU
button without pressing the
S
ELECT
button will allow you to cycle through the menu options.
You must Press the
S
ELECT
button in order to initiate the desired operation.
5ppm
Exit
Zero
Span
mA Span
Alarm1
Alarm2
Alarm3