5.0 Sensor Replacement 

When the sensor module fails bump test or calibration, please replace it with a new one.  
To guarantee the device’s accuracy, it’s suggested to re-calibrate the sensor modules at least once every 6 months.  

 

C

AUTION

:

 

Sensor replacement should only be done by a qualified maintenance  

person. After installing the new sensor module into the device, calibration must be done before using.  

 

6.0 Trouble Shooting Guide 

Normal Problem 

Possible Reasons 

Solution 

Will not power on 

Battery Discharged 

Charge unit  

Battery Defect 

Change Battery 

Circuit fault 

Contact 

ENMET

 / Distributor  

Time display is not correct 

Battery Discharged 

Charge the battery and reset the 
time and date 

Time Zone Difference 

Reset the time and date 

Display “–0” in normal detection 
status 

Sensor drift 

Perform zero calibration 

Zero Calibration function not 
available 

Too much sensor drift 

Recalibrate or replace the sensor 
module 

Reading of gas level not accurate 

Sensor drift 

Recalibrate 

Sensor depleted 

Replace Sensor 

No response to gas 

Warming up not completed 

Wait till it finishes 

Circuit fault 

Contact 

ENMET

 / Distributor 

 

7.0 Technical Data and Specifications 

Electrical Power 

Working Voltage 

3.7 V

DC

, Lithium battery, rechargeable  

 

Operation Time 

Approximate 8 hours, continuous, non-alarm condition 

 

Charge Time 

Approximately 6 hours 

Operation 

Temperature Range 

-20°C to +50°C 

 

Relative Humidity 

5 to 95% RH, non-condensing 

Mechanical 

Dimensions 

2.7 x 4.8 x 1.2 inches (68 x 120 x 30 mm) 

 

Weight 

7.8 ounce (220g) 

 

Explosion Proof 
Ingress Protection 

Exia IICT3 
IP 65 

 

Approval 

Designed to be Intrinsically Safe 

 
N

OTE

:

 

All specifications stated in this manual may change without notice.

 

 

Notes:

 

 

Summary of Contents for RECON/4a

Page 1: ...ENMET 680 Fairfield Court Ann Arbor MI 48108 734 761 1270 Fax 734 761 3220 www enmet com RECON 4a Portable Multi Gas Detector Operation and Maintenance Manual...

Page 2: ...TION 8 5 0 SENSOR REPLACEMENT 10 6 0 TROUBLE SHOOTING GUIDE 10 7 0 TECHNICAL DATA AND SPECIFICATIONS 10 8 0 TERMS AND CONDITIONS 11 8 1 Ordering Information 11 8 2 Delivery 11 8 3 Payment Terms 11 8 4...

Page 3: ...ations Avoid dropping or seriously shocking the instrument NOTE All specifications stated in this manual may change without notice 1 1 Unpack Unpack the RECON 4a and examine it for shipping damage If...

Page 4: ...ng address ENMET 680 Fairfield Court Ann Arbor MI 48108 734 761 1270 Fax 734 761 3220 Toll Free 800 521 2978 1 2 Check Order Check the contents of the shipment against the purchase order Verify that t...

Page 5: ...onal menu functions 3 Sensors Detection area 4 Display The Liquid Crystal Display LCD allows messages to be read clearly 5 Audio Alarm The audio alarm is located on the front of the RECON 4a The audio...

Page 6: ...w Alarm Indication 9 Gas Name 10 Unit of Measure each quadrant 11 Gas Concentration Figure 2 FECON 4a Display 2 3 RECON 4a Pushbutton Function Pushbutton Function Turn on and off the instrument Activa...

Page 7: ...attery Conservation Following the warm up to conserve battery life the backlight will turn off within 30 seconds NOTE Pressing the push button or any gas level alarm reactivates the backlight 3 2 Powe...

Page 8: ...d the display will indicate full charge CAUTION DO NOT start the battery charge process with the RECON 4a power on otherwise the charge speed will be adversely affected Always charge the battery in no...

Page 9: ...ton the RECON 4a will turn OFF Continue to press the push button until the instrument restarts the display indicates CAL and the instrument beeps twice The Display will show Cal followed by a countdow...

Page 10: ...sful wait one hour and repeat the calibration procedure Figure 3 Calibration of RECON 4a 4 1 Calibration Equipment ENMET part numbers Description of Part Part Number Sensor Cover Assembly RECON 4a 026...

Page 11: ...me Zone Difference Reset the time and date Display 0 in normal detection status Sensor drift Perform zero calibration Zero Calibration function not available Too much sensor drift Recalibrate or repla...

Page 12: ...ion by the Seller confirms that the product is defective it will be repaired or replaced at no charge within the stated limitations and returned prepaid to any location in the United States The Seller...

Page 13: ...e Request Form assists in the expedient service and return of your unit and failure to provide this information can result in processing delays ENMET charges a one hour minimum billing for all approve...

Page 14: ...it Appears on Card RETURN SHIPPING METHOD UPS Ground UPS 3 Day Select UPS Next Day Air UPS ND Air Saver UPS 2 Day Air UPS Account FedEx Ground FedEx Air Express Saver FedEx Air Overnight Std FedEx Air...

Reviews: