10
lower right corner will now show “F2”. You will now see
an offset value that is added to the temperature reading
and can be adjusted up or down in the range of -9.0°F to
+9.0°F (-5.0°C to +5.0°C). Adjusting this offset allows you to
fine tune the accuracy of your thermometer in body
temperature measurement mode. This offset does not affect
surface temperature mode readings. Press the RIGHT
button to increase the offset value and the MODE Button to
decrease it. Press the trigger button to confirm the value
you entered.
Enabling/Disabling Sound Output (Beeper)
Long-press the LEFT button until F1 is displayed in the
lower right corner. Press the LEFT button twice briefly, the
lower right corner will now show “F3”. Press the RIGHT
button to toggle the beeper “On” or “Off”. Press the trigger
button to confirm the selection.
Switching Languages
Long-press the MODE button to toggle between English and
Chinese.
Reviewing Stored Body Temperature Readings
When in body temperature measurement mode, the
thermometer automatically stores each reading that was
taken. It can store up to 34 readings, and after 34 readings
have been taken the oldest will be overwritten.