Software Overview and Explanations
Installation and Operation Manual OXITEC
®
5000
20
Doc.-ID: OXI-11022020-EN
4
Software Overview and Explanations
4.1
Menu overview - SYS-MENU
SYS-MENU
System Information
Actual measured values
O
2
measured value (% O
2
) {may be ppm}
O
2
-mA output 17A/B (mA)
O
2
sensor input (mV)
Flow rate reference air (l/h)
O
2
probe temperature (°C / °F)
O
2
probe heater power (%)
Thermocouple input (mV)
Terminal temperature (°C / °F)
Internal temperature (°C / °F)
Process pressure (rel) (mbar/psi)
}
only visible with optional Pressure
O
2
sensor life (%)
Lambda
Calibration results
e.g. 2012-05-11 (Choose date/time)
Executed at
Calibration method
O
2
sensor calibration
~~ Calibration results ~~
O
2
value at test air (20,95 % O
2
)
♦
calibrated to (% O
2
)
O
2
value at test gas (% O
2
)
}
only if determined
♦
calibrated to (% O
2
)
~~ Calibration data ~~
O
2
sensor offset (mV)
O
2
sensor slope (mV / dec)
}
only 2 point calibration
~~ Test gas data ~~
Test air (20,95 % O
2
)
Test gas (e.g. 2,1 % O
2
)
~~ Sensor raw data~~
O
2
voltage at test air (mV)
♦
at pressure (mbar/psi)
O
2
voltage at test gas (mV)
}
only if determined
♦
at pressure (mbar/psi)
O
2
response to test gas (s)
}
only if determined
O
2
response to process (s)
}
only if determined
O
2
sensor life
Device operating data
Powers on counter
Hours in operation
Min. internal temperature
Max. internal temperature