Service and Maintenance
Installation and Operation Manual OXITEC
®
5000
32
Doc.-ID: OXI-11022020-EN
5.6
Replacing the Filter
Warning hot surface
The probe may only be removed with heat-insulated gloves. Before removing the probe, always switch off
the supply voltage to the electronic system. After removal, store the probe in a safe, protected place and
wait until it has cooled down below 35°C/95°F.
Note
Only cemented ceramic, basalt and sintered metal filters can be replaced.
5.6.1
Remove old Filter
Loosen the holding screws.
Clamp the filter head in a vice.
Free the filter by using a hammer
and a metal
extension to force it from the filter
head
Figure 32 -
Remove old filter.
5.6.2
Cement in new Filter
Mix the supplied glue
with water according to the
instructions. Spread the glue into the grooves of the seat
and onto the filter part which comes into contact with the
seat. Push - turn the filter
into the seat. Evenly
smooth the glue between the filter and seat and remove
all excess glue.
Figure 33 -
Cement new filter in
Note
The cement dries at room temperature within 24 hours. Using the cement supplied by ENOTEC, all types
of ceramic, basalt and sintered metal filters supplied by can be inserted and fixated.