Installation and Operation Manual OXITEC
®
5000
System Description
Doc.-ID: OXI-11022020-EN
3
Figure 2 -
OXITEC
®
5000 (probe with cooling tube for flue gas temperatures of up to 1600°C)
Electronic unit SME-5 / IP66
Safe Area - max. ambient temperature:
-20 °C to +55 °C (-4 °F to +131 °F)
InSitu measuring probe / IP65
Safe Area - max. ambient temperature:
-40°C to +80 °C (-4 °F to +176 °F)
Pneumatic cable
Duct / combustion chamber
O
2
probe signal cable
Flue gas direction – max. flue gas temperature 1600 °C
Solenoid valve (optional)
Customer supply
Counter flange (optional)
Manufacturer supply
Isolation: customer
Space required: 2.0 m for standard installation.
0.8 m for 90 °elbow construction
Duct wall
Power supply
Test gas in
Output signals (analog and digital)
Instrument air in