GATEKEEPER GPU OZONE GAS PURIFIERS

3

Installation and Operation Manual  |  Entegris, Inc.

INSTALLATION PROCEDURES

SITE PREPARATION

CAUTION!! Use stainless steel gaskets for fitting 
connection. The use of other gaskets may cause 
abnormal heat generation of the purifier.

1.  Modify the gas line prior to purifier installation using 

a qualified high-purity mechanical contractor.

2. Purge the lines upstream of the purifier to remove 

any atmospheric contaminants.

3. Do not remove the fittings until indicated in this 

instruction manual.

4. Minimize exposure to atmosphere when loosening 

or removing the sealing fittings.

5. Use the proper installation tools and unplated 

stainless steel face seal gasket. If possible, avoid 
using adjustable wrenches, as they often damage 
fittings.

6. Please note, 

never reuse

 face seal gaskets.

INSTALLATION

CAUTION! Take care not to contaminate the 
purifier wetted parts with any foreign substances 
during installation. The contamination may cause 
abnormal heat generation of the purifier.

1.   Verify that the gas line has the proper clearance 

dimensions for the purifier and adjust if needed, 
and all face seals are undamaged.

2.  Orient the purifier in the gas line with the flow 

arrow pointing in the direction of the gas flow.

3.  Prepare two new stainless steel face seal gaskets 

before removing purifier caps. Be sure to work 
quickly, completing the next five steps without 
interruption. This will minimize the purifiers 
exposure to atmosphere. 

4. Inspect inlet and outlet face seals for damage.

5.  Insert a new metal seal gasket for the outlet and 

hand tighten the connection.

6.  Insert a new metal seal gasket for the inlet and 

hand tighten the connection.

NOTE: Take care to not rotate the purifier body. This 
may damage the sealing surface, which could lead to  
a leak.

7.   Using the proper wrenches, tighten the inlet and 

then the outlet connections an additional 1⁄8 of a 
turn (45°) past finger tight to seal the connection. 
The purifier must be held stationary while turning 
the mating fitting. Do not over tighten as this may 
cause damage to the fittings.

8.  Perform the leak test according to procedure that is 

defined at site.

BEFORE STARTUP

Ensure the purifier has been leak checked and 
installed as recommended. Also ensure the purifier  
is being operated within its operating limits, such as 
temperature and pressure.

FLOWING OZONE GAS

CAUTION! Monitor the purifier surface tem- 
perature when starting service in ozone gas.  
If the purifier is contaminated with any foreign 
substances during installation, the purifier may 
experience abnormal heat generation.

1.  Flow a minimum of 20 liters of ozone gas in the 

purifier inlet through the outlet for a minimum of 
five minutes.

2. Ensure the purifier surface temperature is lower 

than 40°C (104°F) while ozone gas is flowing. If  
the surface temperature exceeds 40°C (104°F), 
immediately turn off ozone gas flow and remove 
the purifier according to the following decommis-
sioning procedure.

3. Purifier is now ready for use.

Summary of Contents for GateKeeper

Page 1: ...MICROCONTAMINATION CONTROL MANUAL GateKeeper GPU Ozone Gas Purifiers Installation and operation manual ...

Page 2: ...allation Procedures 3 Site Preparation 3 Installation 3 Before Startup 3 Flowing Ozone Gas 3 Decommissioning Procedure When Heated 4 Managing Replacement and Purifier Lifetime 4 Disposal 4 Troubleshooting Guide 5 Mechanical Specifications 6 GPU Options 6 Limited Warranty 7 For More Information 7 Terms and Conditions of Sale 7 ...

Page 3: ...t and or mitigate hazardous conditions Ensure the area is well ventilated to prevent asphyxiation DANGER USE THIS PURIFIER WITH APPROVED GASES ONLY If this product is exposed to an incompatible gas an exothermic or other adverse reaction may occur It is the user s responsibility to install all safety equipment to prevent and mitigate the effects of such a condition WARNING Always pressurize and de...

Page 4: ...hout interruption This will minimize the purifiers exposure to atmosphere 4 Inspect inlet and outlet face seals for damage 5 Insert a new metal seal gasket for the outlet and hand tighten the connection 6 Insert a new metal seal gasket for the inlet and hand tighten the connection NOTE Take care to not rotate the purifier body This may damage the sealing surface which could lead to a leak 7 Using ...

Page 5: ...e operating site 5 Cap the purifier tightly with metal caps CAUTION DO NOT REUSE the decommissioned purifier MANAGING REPLACEMENT AND PURIFIER LIFETIME Since the series does not have an end of life indicator it is recommended that they be put on a Preventative Maintenance PM cycle based on the estimated lifetime In order to determine the purifier s lifetime please visit www entegris com or contact...

Page 6: ...nstream parts generate metals after purification Install the purifier near the use point to avoid metal generation by downstream parts Purifier flow rate is too high Verify that the maximum flow rate is not exceeded Outlet purity level degrades with time Purifier has exceeded its useful life Replace the purifier with a new purifier Ozone concentration declines after purification The inertness of t...

Page 7: ...al removal efficiency 99 for Cr Particles 99 9999999 for particles 0 003 µm Connection 1 4 gasket seal Gas Ozone O 300 g Nm3 concentration Inlet metal level Cr 1 ppm Recommended flow rate 5 to 25 slm Operating conditions Maximum operating pressure 20 kPa 2 9 psi Operating temperature 20 to 30 C 68 to 86 F Please consult Entegris Account Managers for flow rate and O gas concentration exceeding spec...

Page 8: ...it entegris com and select the Contact Us link to find the customer service center nearest you TERMS AND CONDITIONS OF SALE All purchases are subject to Entegris Terms and Conditions of Sale To view and print this information visit entegris com and select the Terms Conditions link in the footer www entegris com Installation and Operation Manual Entegris the Entegris Rings Design and other product ...

Reviews: