ENTEGRIS, INC.
Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300
13
NT® PRESSURE T RAN SDUCER USE R GUIDE
NT® PRESSURE TRANSDUCER USER GUIDE
12
Customer Service Tel. 763-502-0200 Customer Service Fax 763-502-0300
ENTEGRIS, INC.
Troubleshooting the NT
®
Pressure
Transducers may be accomplished
by measuring the output signal of
the pressure transducer with a
battery powered current/voltage
meter. The meter may be placed in
series with the pressure transducer
to measure the current output or it
may be used to directly measure the
voltage output.
Using the battery powered current/
volt meter is an effective method to
determine whether the device or the
on-site data acquisition system is not
functioning properly.
Troubleshooting
The following lists the specifications
for NT
®
Pressure Transducers,
models 4150 and 4250. Consult the
factory for product specifications
manufactured for nonstandard
applications.
NOTE:
Specifications are subject
to change without notice. Please
consult the factory for the most
current information.
Reference
CONSTRUCTION MATERIALS
Wetted parts:
Body:
PTFE
Sensor Interface:
Sapphire
O-Ring:
Kalrez
®
4079, Kalrez
®
1050 or Kalrez® 6375 UP
Nonwetted parts:
Polypropylene, polyethylene and PVDF
(In addition to materials listed above)
Connection type:
Flaretek
®
Physical Specifications
Electrical Specifications
Input power:
24 VDC (12-28 VDC)
Pressure signal output ranges:
4-20 mA
Electrical connection:
6
′
, 12
′
, 30
′
FEP-coated pigtail or polypropylene
3-pin connector with PVC cable
Reference accuracy:
±1% of span (includes linearity, hysteresis
and repeatability) at 73°F (23°C)
Process temperature:
50° to 95°F (10° to 35°C)
Performance Specifications