20
OPTIMIZER ST2 FILTER MANIFOLD
Installation and Use Manual | Entegris, Inc.
5. Ensure the O-rings are not tilted or distorted. If they
are, please follow the Replacing O-ring procedure
on page 21.
CAUTION: When removing or attaching the filter
or flushing shell, do not touch anywhere other
than the lever. The shaft may come off and an
unexpected accident such as leakage may occur.
There is also a safety risk of getting your fingers
caught in moving parts, resulting in injury.
6. Lower the lever until you hear a click from the
release tab.
CAUTION: Make sure all the spring shafts are
properly engaged. See page 14 for detailed check
points. Make sure all the fittings and tubing are
securely fastened before starting liquid flow.
FILTER STARTUP
—
1. Connect a suitable filter using the installation
procedure described in this manual.
2. Start the liquid flow when the dispense pump or
pressurized system is ready.
3. Open the valve on the filter vent line until liquid
comes out and exhaust the air inside the filter and
the pipelines.