Setting Up and Collecting Off Route Time Waveform Measurements
dataPAC User’s Guide
129
5.
Use the up and down arrow keys to select Spectra / Waveform Measurement
Parameters, then press <SELECT>. The dataPAC displays the setup parameters
window for the selected measurement type.
Note:
If the Spectra Parameters screen appears, press <F5> (TWF Parameters) to get to the
Time Waveform screen.
6.
Press <F1> (Edit Point) to change the measurement setup. You can press
<SHIFT><F1> (Point Context) to copy the current point setup to your off route
measurement. Alternatively, you can press <SHIFT><F2> (Instrument Defaults) to
select the instrument defaults for your off route measurement. Use the field descriptions
below to set up the measurement.
Xdcr Native Units
The Transducer Native Units parameter allows you to specify the type of transducer
connected to your dataPAC. The choice of this parameter controls the list of possible
selections available for the Measurement Variable parameter.
Meas. Variable
The Measurement Variable parameter allows you to choose how to measure the signal
coming from the transducer. For example, if you have a velocity probe, you can measure
either velocity or displacement. The Measurement Variable parameter also includes gSE for
the dataPAC 1500.
Unit Text
The Unit Text parameter allows you to select the display units for this measurement. Select
from RMS, Pk, or Pk-Pk.
The list of selections is dependent upon current selections in the previous two parameters:
Transducer Native Units and Measurement Variable. Note that this simply specifies the text
that is used to label the measurement and scales the data appropriately.
Hardware Range
The Hardware Range parameter is used to set a maximum range for this measurement.
Selecting Auto Range causes the system to select an optimum range after you press
<STORE> to start a measurement. Generally, it is recommended that you use Auto Range.
However, if want to specify a range, you can select one from the list.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com