Setting Up and Collecting Off Route Phase/Magnitude/Speed
dataPAC User’s Guide
141
Trigger Control
This parameter controls the slope type used for the trigger input. Your selections are:
z
Positive Slope - Ext - Defines the trigger event to occur on a rising edge.
z
Negative Slope - Ext - Defines the trigger event to occur on a falling edge.
z
Leading Edge/Trailing Edge - Use this parameter to have the dataPAC 1500
automatically select the rising edge or the falling edge of the trigger mark. The
selection is dependent upon the trigger pulse width and polarity. For example, if the
trigger input is a narrow positive pulse, then Leading Edge selects the rising edge of the
trigger event and Trailing Edge selects the falling edge.
z
Positive Slope - Int - This parameter is automatically selected if you choose Strobe for
the Trigger Select parameter.
Collecting Off Route Phase/Magnitude/Speed Measurements
1.
Press <STORE> to start the measurement. The dataPAC ranges and collects the data.
2.
Once the dataPAC collects the measurement, press <STORE> again to store the
measurement. You can view the measurement in progress for as long as you wish
without storing. If in minimum mode, the dataPAC displays a small folder with the
number 1 in it after you press <STORE>. This indicates that the dataPAC has saved the
measurement. If you have Data Overwrite enabled, only the data from the last time you
pressed <STORE> is unloaded to your database.
Note:
See “Selecting the Data Collection Options” on page 127 for more information about
the Data Overwrite option.
The measurement is stored as an off route measurement at the point displayed in the
reduced view window.
3.
Press <DONE> when you are finished. You can view the last measurement, if you
stored it, by pressing <F3> (Review Data).
Storage point