MN-28646 • Rev 4 • 03/18 • EC Series Fluid Combination Blanket/Fluid Warmer
8
Electrical Information
Hazardous
Voltage Present
IMPORTANT
Do not load the metal basket beyond the recommended
maximum capacity:
EC1260BL = 18 1-liter bottles/basket or
28 1-liter bags/basket
EC1540BL = 18 1-liter bottles/basket (54 total) or
14 1-liter bags/basket (42 total)
EC1730BL = 30 1-liter bottles/basket or
24 1-liter bags/basket
Overloading may cause lower or uneven temperatures of
product and damage to basket and basket rail supports.
Baskets that are overloaded may slip off rail supports,
resulting in possible damage to product and equipment,
as well as causing possible injury.
CAUTION
Power source must match voltage identified
on appliance rating tag. The rating tag
provides essential technical information
required for any appliance installation,
maintenance or repairs. Do not remove,
damage or modify the rating tag.
DANGER
To prevent serious personal injury, death,
or property damage:
Do not use this warming appliance in
the presence of flammable anesthetic
mixtures (with air or with oxygen or
nitrous oxide).
Not category AP or APG equipment
CAUTION
This appliance is not approved for
warming blood or blood products.
WARNING
Injection fluid manufacturer suggests not to warm
injection fluids above 40°C (104°F). If fluids are warmed
above the suggested temperature, the fluids should
be discarded.
120 V.A.C. — 50/60 Hz, 1 ph
1.7 kW, 14.2 Amps
Safety Class I Equipment
No Applied Parts
Mode of Operation: Continuous
NEMA 5-20P
20A - 125V Plug
Hospital Grade
IP X0
EC1730BL POWER REQUIREMENTS
EC1260BL POWER REQUIREMENTS
120 V.A.C. — 60 Hz, 1 ph
1.92 kW, 16.0 Amps
Safety Class I Equipment
No Applied Parts
Mode of Operation: Continuous
NEMA 5-20P
20A - 125V Plug
Hospital Grade
IP X0
120 V.A.C. — 50/60 Hz, 1 ph
1.92 kW, 16.0 Amps
Safety Class I Equipment
No Applied Parts
Mode of Operation: Continuous
NEMA 5-20P
20A - 125V Plug
Hospital Grade
IP X0
EC1540BL POWER REQUIREMENTS
Safety Class I
Equipment
Protective Earth
Ground Symbol
Medical Equipment classified by Underwriters
Laboratories with Respect to Electrical Shock, Fire
and Mechanical Hazards only, in Accordance with
UL 60601-1 and CAN/CSA C22.2 No. 601.1.
UL File No.
E201645
Grounding reliability can only be achieved when equipment is
connected to an equivalent receptacle marked “Hospital Grade.”
Wire diagram located under top cover of unit
Locate the Rating Tag
Verify the power requirements for the appliance. The power specifi cation is located on the
appliance identifi cation rating tag. This tag is permanently attached to the appliance.
Hazardous
Voltage Present
To prevent an electrical shock hazard between the appliance and
other appliances or metal parts in close vicinity, an equalization-
bonding stud is provided. An equalization bonding lead must be
connected to this stud and the other appliances / metal parts
to provide sufficient protection against potential difference.
The terminal is marked with the following symbol.
*Other international plugs are available, contact factory for more information.