background image

MN-28646 • Rev 4 • 03/18 • EC Series Fluid Combination Blanket/Fluid Warmer

4

General Information

This warming cabinet is designed to safely warm and store 

blankets and either irrigation fl uids or injection fl uids.
The dual-chambered warming cabinet is constructed with 

stainless steel exterior casing and door with handle and hinges 

designed to withstand heavy usage. A door with window 

allows observation of inventory with the door closed. The 

cabinet is warmed using low-heat-density electrothermal 

cable array. The electrothermal cable is positioned in the fl oor 

and two sides of the warming cabinet, providing even heating 

of the interior chamber.
Each chamber's temperature is regulated by a separate 

electronic control consisting of a 4 digit L.E.D. display, 

ON/OFF button, INCREASE and DECREASE buttons, 

integrated lock-out control feature and a series of prompt 

sequence indicators.
Each control can easily be set to operate in Fahrenheit or 

Celsius. Aft er a power failure, the cabinet will remember its 

programming and begin to operate as before. The ON/OFF 

indicator will blink to indicate a failure occurred; pressing the 

ON/OFF button once will eliminate this blinking. A thermal 

shut-off  system, separate from the electronic controls, is 

included as an additional safety feature.
The control will display temperature in whole degrees.

Blanket warming chamber:
The design and operational characteristics of the chamber 

eliminate the need for a heat circulating fan.
The controller has an adjustable temperature range of 

37°C to 93°C (98°F to 200°F).
The timer feature allows the user to program the controller 

to automatically switch on and switch off  once during a 

24-hour period at selected times. This enables the chamber to 

be turned off  automatically at night to save energy, but turned 

on again in the early morning to ensure warm blankets are 

available. 

Fluid warming chamber:
The warming appliance can be programmed to warm 

either irrigation fl uids (IRR) or injection fl uids (INJ), with 

separate temperature ranges provided depending on the 

choice selected. 

•  IRR temperature range: 37°C to 66°C (98°F to 150°F), 
•  INJ temperature range: 37°C to 40°C (98°F to 104°F). 

A fan located inside the chamber mixes the air to prevent 

temperature stratifi cation and to ensure an accurate 

chamber temperature for each mode. 

•  0/-1.12°C (+0/-2ºF) for set points of 

37°C to 40°C (98°F to 104°F).

•  0/-1.67°C (+0/-3ºF) for set points of 

43°C to 66°C (110°F to 150°F)

An alarm will sound if temperatures exceed 6°C (10°F) over 

the set-point temperature, and an overtemp indicator will 

blink indicating an over-temperature condition. 

The blanket warming chamber includes a white, epoxy-

coated blanket support assembly and two (2) shelves. The 

fl uid warming chamber is equipped with one (1) white, 

epoxy-coated shelf to accommodate most commonly used 

surgical fl uids packaged in bags or bottles. The basket has a 

capacity of 

18 1-liter bottles or 28 1-liter bags.

 The cabinet is 

furnished with four (4) leveling feet.

EC1260BL INFORMATION:

The blanket warming chamber contains an epoxy-coated 

blanket support assembly and one (1) shelf. The fl uid 

warming chamber is equipped with three (3) white, epoxy-

coated wire baskets to accommodate most commonly used 

surgical fl uids packaged in bags or bottles, mounted on 

basket rail supports. Each basket has a capacity of 18 1-liter 

bottles or 24 1-liter bags. The cabinet is furnished with a full 

perimeter rubber bumper assembly for exterior protection, 

and one (1) set of 5" (127mm) heavy-duty casters – two (2)

rigid and two (2) swivel with locking brake.

EC1540BL INFORMATION:

The blanket warming chamber contains an epoxy-coated 

blanket support assembly and two (2) shelves. The fl uid 

warming chamber is equipped with one (1) white, epoxy-

coated metal basket to accommodate most commonly used 

surgical fl uids packaged in bags or bottles, mounted on 

basket rail supports. The basket has a capacity of 30 1-liter 

bottles or 24 1-liter bags. The cabinet is furnished with one 

(1) set of 5" (127mm) heavy-duty casters – two (2) rigid and 

two (2) swivel with locking brake.

EC1730BL INFORMATION:

    WARNING

To prevent serious personal injury, death, or 

property damage: 
Do not steam clean, hose down or flood the 

interior or exterior with water or liquid solution 

of any kind. 

Do not use water jet to clean. Failure 

to observe this precaution will void the warranty.

(IP X0 - Listed as Ordinary)

DANGER

To prevent serious personal injury, death, 

or property damage:

Do not use this warming appliance in 

the presence of flammable anesthetic 

mixtures (with air or with oxygen or 

nitrous oxide). 

Not category AP or APG equipment 

Appliance Description and Intended Use

Summary of Contents for EC1260BL

Page 1: ...Systems An ISO 13485 2003 certified company W164 N9221 Water St Menomonee Falls WI 53051 Tel 262 251 8356 800 TO B WARM generalinfo enthermics com www enthermics com Printed in the U S A Specification...

Page 2: ...Company www enthermics com Delivery 1 Transportation Damage and Claims 1 Unpacking 2 Safety Procedures 3 General Information 4 Dimensions 5 Electrical Information 8 Operation Instructions 9 Cleaning a...

Page 3: ...package has been damaged or opened All Enthermics Medical Systems appliances are sold Free on Board F O B shipping point and when accepted by the carrier such shipments become the property of the cons...

Page 4: ...g operators Additional manuals are available from the service department 2 Read all instructions in this manual carefully before initiating the installation of this appliance using the appliance or pe...

Page 5: ...ncrease patient comfort The fluid chamber is only intended for warming medical solutions for irrigation and injection prior to use Refer to the labeling of the manufacturer of the products to be warme...

Page 6: ...prevent temperature stratification and to ensure an accurate chamber temperature for each mode Within 0 1 12 C 0 2 F for set points of 37 C to 40 C 98 F to 104 F Within 0 1 67 C 0 3 F for set points o...

Page 7: ...NGTH 8 2 438mm est 4 10 104 0mm CORD 1 7 43 0mm CORD 50 1270mm 29 8 756mm RADIUS 30 1 764mm 12 5 316mm CAVITY 32 6 827mm CAVITY 26 6 674mm CAVITY 30 9 785mm WITH FEET 0 6 15mm FEET 66 6 1692mm 69 2 17...

Page 8: ...8 248mm SHELF 23 7 603mm INSERT 72 7 1846mm 18 7 476mm CAVITY 25 2 639mm 5 5 141mm 26 6 676mm CAVITY 26 9 683mm CAVITY 11 2 285mm 54 6 1387mm 28 9 735mm 35 1 891mm CORD LENGTH 8 2 438mm est 33 1 841m...

Page 9: ...2 5 826mm CAVITY 23 3 592mm CAVITY 28 8 732mm 73 4 1864 mm 26 7 678mm CAVITY 31 1 790mm ON OFF LOCK TIME START STOP TIMER ON OFF OVER TEMP IRR INJ LOCK OVERTEMP ERROR 27 6 700mm INSERT 4 6 117mm 4 0 1...

Page 10: ...ps Safety Class I Equipment No Applied Parts Mode of Operation Continuous NEMA 5 20P 20A 125V Plug Hospital Grade IP X0 EC1730BL POWER REQUIREMENTS EC1260BL POWER REQUIREMENTS 120 V A C 60 Hz 1 ph 1 9...

Page 11: ...w button 5 simultaneously The temperature at the compartment sensor is displayed 8 How to Lock the Controller The controller can be locked to prevent changes from being made to the temperature set poi...

Page 12: ...ary press the up arrow button 5 or the down arrow button 10 to change the start time 3 Press the TIMER button 11 The STOP icon 4 illuminates This is the time of day the controller automatically shuts...

Page 13: ...receptacle specified on the electrical information page 2 Locate the circuit breaker at the back of the warmer 3 Push the circuit breaker to the ON I position 4 Press the ON OFF button 5 The on off s...

Page 14: ...Overtemp When the electronic control senses a temperature 6 C 10 F greater than the set point this indicator illuminates the audible alarm sounds and the on off button indicator blinks The green indi...

Page 15: ...the IRR button to select the irrigation fluids mode or the INJ button to choose the injection fluids mode The last set point temperature for that mode of operation will appear in the display 5 Set des...

Page 16: ...o avoid surface scratches Never use wire brushes metal scouring pads or scrapers to remove residue Failure to observe this will void the warranty CAUTION To protect surfaces completely avoid the use o...

Page 17: ...ng solution after cleaning and rinsing This solution must be approved for use on stainless steel surfaces Replace support assembly 5 Clean the exterior of the cabinet with a cleaner recommended for st...

Page 18: ...exceed the height of the insert Measure the resistance of the sensor with an ohmmeter Test the sensor at 32 F 0 C using a container of ice water The ohm reading should be 100 If the ohm reading is 10...

Page 19: ...er than set point Cavity temperature is higher than the set point temperature Allow cavity to cool to set point temperature Not applicable on warmers with control CC 34765 door The door has been open...

Page 20: ...IGID 2 CS 24874 2 CS 24874 24 CASTERS 5 127mm SWIVEL W BRAKE 2 CS 24875 2 CS 24875 25 FAN SWITCH 1 SW 33907 26 SCREWS FAN SWITCH 8 32 x 1 FH PHH 2 SC 22138 27 SCREWS FAN SWITCH BRACKET 8 32 x 1 4 FH P...

Page 21: ...tion Blanket Fluid Warmer 19 Service ON OF F LO CK TIM E STAR T STOP TIM ER ON OF F OV ER TEMP IRR INJ LO CK OV ER TEMP ER RO R 1 2 3 4 9 40 11 17 15 16 22 43 35 45 34 33 36 1 12 13 14 20 8 4 5 10 9 3...

Page 22: ...lanket Fluid Warmer 20 Service ON OFF LOCK TIME START STOP TIMER ON OFF OVER TEMP IRR INJ LOCK OVERT EMP ERROR 17 14 20 25 28 27 33 35 21 23 24 32 18 26 34 12 13 19 1 8 4 5 10 9 3 7 6 11 15 16 2 41 40...

Page 23: ...tion Blanket Fluid Warmer 21 Service ON OFF LOCK TIME STAR T STOP TIMER ON OFF OVER TEMP IRR INJ LOCK OVER TEMP ERRO R 33 11 14 15 36 34 23 24 1 1 2 3 4 9 8 4 5 10 9 3 7 6 35 32 37 21 20 17 16 12 13 4...

Page 24: ...1 TT 34600 9 COVER HI LIMIT 2 1009751 2 1009751 2 1009751 10 POWER SUPPLY BOARD 1 BA 34596 1 BA 34596 1 BA 34596 11 RELAY 12V DC COIL 3 RL 34434 2 RL 34434 3 RL 34434 12 TERMINAL BLOCK PORCELAIN 1 BK...

Page 25: ...6 Rev 4 03 18 EC Series Fluid Combination Blanket Fluid Warmer 23 Service 3 9 7 11 16 11 2 8 14 15 13 5 6 10 16 1 2 3 7 9 8 5 6 11 10 4 13 14 15 13 14 17 17 EC1540BL Electrical View EC1730BL Electrica...

Page 26: ...2m SL 3063 INSULATING SLEEVE 6 12 6 CR 3226 RING CONNECTOR 6 12 6 BU 3105 SHOULDER BUSHING 6 12 6 ST 2439 STUD 10 32 6 12 6 NOT SHOWN ACCESSORY SERVICE PARTS EC1260BL EC1540BL EC1730BL Built in Trim K...

Page 27: ...8 EC Series Fluid Combination Blanket Fluid Warmer 25 Serial number is required for all inquiries Part numbers and drawings are subject to change without notice Wire Diagrams Refer to wire diagram inc...

Page 28: ...8 EC Series Fluid Combination Blanket Fluid Warmer 26 Serial number is required for all inquiries Part numbers and drawings are subject to change without notice Wire Diagrams Refer to wire diagram inc...

Page 29: ...8 EC Series Fluid Combination Blanket Fluid Warmer 27 Serial number is required for all inquiries Part numbers and drawings are subject to change without notice Wire Diagrams Refer to wire diagram inc...

Page 30: ...5 2003 certified company W164 N9221 Water St Menomonee Falls WI 53051 Tel 262 251 8356 800 TO B WARM generalinfo enthermics com www enthermics com Printed in the U S A Specifications are subject to ch...

Reviews: