ENTHERMICS MEDICAL SYSTEMS
O C M / E C 1 5 4 0
bl
0 1 / 0 7
IMPORTANT
Do not load the metal basket beyond the
recommended maximum capacity of
14 liters
.
Overloading may cause lower or uneven
temperatures of product and damage to basket and
basket rail supports. Baskets that are overloaded
may slip off rail supports, resulting in possible
damage to product and equipment, as well as
causing possible injury.
E L E C T R I C A L I N F O R M A T I O N
The power specifications are located on the unit identification nameplate. This nameplate is
permanently attached to the unit and must be located to verify power requirements.
P R E P A R A T I O N
Before operating the cabinet, clean both the interior and exterior of the unit with a damp cloth and mild
soap solution. Wipe with an appropriate disinfectant. Clean and install the cabinet basket support
assembly and the blanket support assembly.
Hazardous
Voltage Present
DO NOT use this warming cabinet in the
presence of a flammable anethestic mixture
(with air or with oxygen or nitrous oxide).
THIS COULD RISK AN EXPLOSION!
INJECTION FLUIDS SHOULD NOT
BE WARMED OVER 110
°
F (43
°
C)
THIS UNIT HAS NOT BEEN
APPROVED FOR WARMING OF
BLOOD OR BLOOD PRODUCTS.
To prevent an electrical shock hazard between the appliance and other
appliances or metal parts in close vicinity, an equalization-bonding stud
is provided. An equalization bonding lead must be connected
to this stud and the other appliances / metal parts to provide
sufficient protection against potential difference. The terminal
is marked with the following symbol.
Protective Earth
Ground Symbol
POWER REQUIREMENTS - EC1540
bl
(
120V)
120 V.A.C. — 50/60 Hz, 1 ph
1920 Watts, 16.0. Amps
Safety Class I Equipment
Medical Equipment classified by Underwriters
Laboratories with Respect to Electrical Shock,
Fire and Mechanical Hazards only, in Accordance
with UL 60601-1 and CAN/CSA C22.2 No. 601.1.
UL File No.
E201645
Grounding reliability can only be achieved when equipment is
connected to an equivalent receptacle marked "Hospital Grade."
NEMA 5-20P
20A - 125V Plug
Hospital Grade
POWER REQUIREMENTS - EC1540
bl
(
230V)
230 V.A.C. — 50/60 Hz, 1 ph
1700 Watts, 7.5 Amps
Type B Equipment
BS 1363 Plug
(UK only)
CEE 7/7
220-230V Plug