7
ENTHERMICS MEDICAL SYSTEMS
O C M / E C 1 5 4 0
bl
0 1 / 0 7
AUTOMATIC 24-HOUR TIMER PROGRAMMING
With the control off,
press and hold
the TIMER key to
illuminate various indicators. Release. Repeat the process
until the Start status indicator is illuminated.
Continue to
hold
the TIMER key and set desired start time by using the
UP or DOWN arrow keys. Release the TIMER key. Press
and hold the TIMER key again. The STOP status indicator
will illuminate. Set the desired stop time by using the UP
or DOWN arrow keys. Release the TIMER key. The
TIMER key status indicator will illuminate. The auto timer
mode is programmed. To change any of these settings, use
the TIMER key and the UP or DOWN arrow keys.
Note: Each press of the TIMER key will display one of
the settings: start time, stop time, or current time of day;
each of which can be adjusted using the UP or DOWN
arrow keys.
The compartment will start heating at the set
start time with the ON/OFF status indicator L.E.D.
illuminated. At the preset stop time the unit will
discontinue heating.If none of the parameters are changed,
the unit will start and stop heating at the same times the
next day. The TIMER status indicator is illuminated, if the
control is in the auto timer mode. To turn off the timer,
press and hold
the TIMER key until the TIMER key status
indicator is extinguished.
CONTROL LOCK PROGRAMMING
The warmer control can be locked so that no changes can
be made to the temperature set-point or the mode selection.
Press and hold
the ON/OFF key and the UP arrow key at the
same time. The unit will beep and the LOCK indicator will
illuminate. Attempts to operate the ON/OFF key, or to
change the temperature set-point will be unsuccessful. To
unlock the warmer control,
press and hold
the ON/OFF key
and the DOWN arrow button at the same time. The unit
will beep, the control will unlock, and the LOCK indicator
will extinguish.
ON
OFF
LOCK
TIME
START
STOP
L.E.D.
DISPLAY
STATUS INDICATOR L.E.D.
UP ARROW
DOWN ARROW AUTO TIMER
KEY
ON / OFF
KEY
L.E.D. = LIGHT EMITTING DIODE
TIMER
THERMOSTAT CONTROL AND L.E.D. DISPLAY
1. The appliance should be plugged into a hospital grade,
NEMA 5-20P receptacle
(120V) or an appropriate
receptacle for a 230V appliance.
2. Turn on the power circuit breaker
switch, which is located at the back
of the appliance. It is a rocker-type
switch with international ON (
I
) and
OFF (O) markings.
3. ACTIVATE CONTROL BY PRESSING
THE ON/OFF KEY ON CONTROL PANEL ONCE.
The ON/OFF indicator will illuminate and remain lit
until the unit is turned off. The digital L.E.D. display
will indicate last temperature set-point of compartment.
4. SET DESIRED TEMPERATURE.
To set the blanket warming temperature,
press and hold
the UP or DOWN ARROW keys to change the value
shown in the display. The temperature set-point range
is 90° to 200°F (32° to 94°C).
5. LOAD THE UPPER CHAMBER WITH BLANKETS.
Check that the epoxy-coated blanket support assembly
is in place. A full load of blankets will take 2-3 hours to
reach optimum temperature. Make certain the cabinet
door is securely closed after initial loading and following
each blanket removal.
Note:
Do not block sensor by overloading cabinet
with blankets.
B L A N K E T C O N T R O L F E A T U R E S
( U P P E R C H A M B E R )
B L A N K E T O P E R A T I O N A L P R O C E D U R E S
( U P P E R C H A M B E R )
POWER
CIRCUIT BREAKER
SWITCH
Blanket Warming Chamber