MN-28671
Rev 2 • 230V Enthermics Fluid Warming Cabinet Operation & Care Manual • 8
1.
The appliance should be plugged
into an appropriate
hospital grade receptacle for a 230V appliance.
2. Turn on the power circuit breaker
switch, which is located at the back
of the appliance. It is a rocker-type
switch with international ON (
I
) and
OFF (O) markings.
3. ACTIVATE CONTROL BY PRESSING THE
ON/OFF KEY ON CONTROL PANEL ONCE.
The ON/OFF indicator will illuminate and remain lit
until the unit is turned off. The digital display will
indicate last temperature set-point of compartment.
4. SELECT DESIRED MODE OF OPERATION.
Press the IRR key to select the IRRIGATION FLUIDS
mode or the INJ key to choose the INJECTION FLUIDS
mode. The last set-point temperature for that mode of
operation will appear in the display.
NOTE:
In order to switch between the irrigation and
injection modes, you must first turn the unit
off and then turn the unit back on.
5. SET DESIRED TEMPERATURE.
To set the fluid warming temperature,
press and hold
the
UP or DOWN ARROW keys to change the value shown
in the display. The IRRIGATION FLUIDS set-point
temperature range is 37° to 66°C (98° to 150°F) and
the INJECTION FLUIDS set-point temperature range
is 37° to 40°C (98° to 104°F).
NOTE:
The warmer is designed to warm fluids to the
appropriate temperature recommended by your
supplier. The warm-up stabilization time will vary
depending on the warmer load. Exercise judgment
to determine inventory rotation protocols and
warm-up time for the fluids you use.
Caution:
• Check fluid temperature prior to use.
• Verify that the fan at the back or top of the chamber is
rotating freely. If it is not working, discard your inventory,
contact your service representative, and discontinue use of
unit until it is repaired.
• If the warmer control has failed, or if error messages
are displayed, discard your inventory and contact your
service representative.
• The unit may tip over if more than one drawer is
extended simultaneously.
Open only one drawer at a time
when loading or unloading fluids.
POWER
CIRCUIT BREAKER
SWITCH
O P E R A T I O N A L P R O C E D U R E S
PO
WER
CIRCUIT BREAKER
SWITCH
or
W A R N I N G
REFER TO FLUID MANUFACTURER’S
LABELING FOR RECOMMENDED
WARMING PROCEDURES
CAUTION
INJECTION FLUID MANUFACTURER
SUGGESTS NOT TO WARM INJECTION
FLUIDS OVER 40°C (104°F).
W A R N I N G
TRANSPORT SHALL ONLY BE DONE
WITH THE DOORS CLOSED
D A N G E R
DISCONNECT UNIT FROM
POWER SOURCE BEFORE
CLEANING OR SERVICING.
D A N G E R
DO NOT use this warming appliance
in the presence of flammable
anesthetic mixture (with air or
with oxygen or nitrous oxide).
THIS COULD RISK AN EXPLOSION!
(Not category AP or APG equipment )