background image

 

- 8 - 

4.2.2

 

Calibration mode 

The meter can adopt 1 to 3 points of calibration. The 1

st

 point must be using 7.00 pH (or 6.86 pH) 

buffer solution, and then choose other buffers to do 2

nd

 point and 3

rd

 point. Please refer to Table-3 

for details.

 

Table - 3 Calibration mode 

                                                                                       

 

 

USA

 

NIST

 

Calibration 

icon

 

When to adopt

 

1-Point 

Calibration

 

7.00 pH

 

6.86 pH

 

         

 

Accuracy: ±0.1pH

 

2-Point 

Calibration

 

7.00 pH→4.00 or 1.68pH

 

6.86 pH→4.01 or 1.68pH

 

 

 

Measuring range: 0 

to 7.00 pH

 

7.00 pH→10.01 or 12.45pH

 

6.86 pH→9.18 or 12.45pH

 

           

 

Measuring 

range: >7.00 pH

 

3-Point 

Calibration

 

7.00pH→4.00 or 

1.68pH→10.01 or 12.45pH

 

6.86pH→4.01 or 1.68pH→

 

9.18 or 12.45pH

 

 

Wide measuring 

range

 

 

4.3 

pH Measurement

 

4.3.1

 

The 1

st

 point calibration 

Rinse the pH electrode and the temperature probe in distilled or deionized water and allow it to dry, 
long press          to the enter calibration mode. CAL1 icon will flash in the bottom right corner of the 
LCD, reminding you to use pH7.00 buffer to conduct the 1

st

 point of calibration. Dip the pH electrode 

and the temperature probe into pH7.00 buffer solution, stir gently and let it stand still and wait for 
the reading to be stable. When the stable icon          displays on LCD, press                to finish 1

st

 point 

calibration, meter goes to measurement mode, the bottom left of LCD indicating        . 

 

 

4.3.2

 

The 2

nd

 point calibration 

Rinse  the  electrodes  in  distilled  or  deionized  water,  allow  it  to  dry.  Long  press          to  enter 

calibration mode. CAL2 icon flashes at the bottom right corner of the LCD, reminding you to make 

the 2

nd

 point of calibration. Submerge the electrodes into pH4.00 buffer solution, stir the solution 

gently and allow it to stay in the buffer solution until a stable reading is reached. The meter’s display 

will  show  scanning  and  locking  process  of  calibration  buffer  solution  at  the  bottom  right  of  LCD. 

When the meter locks 4.00pH, stable icon          displays and stays on LCD. Press                to calibrate 

the meter. Electrode slope displays after calibration is done, the 2

nd

 point calibration finished, the 

meter goes to measurement mode, and the bottom left of LCD indicates              .

 

Summary of Contents for 35613-65

Page 1: ...ION 100 Portable Ion Meter Instruction Manual 35613 65 68 Environmental Express www environmentalexpress com...

Page 2: ...nformation regarding pH calibration and measurement 9 4 6 Customized calibration take 1 60pH 6 50pH as example 10 4 7 Notes regarding customized calibration 10 4 8 Self diagnosis information 10 4 9 Ma...

Page 3: ...der to use and maintain the instrument properly please read the manual thoroughly before use To improve instrument performance constantly we reserve the right to change the manual and accessories with...

Page 4: ...86 9 18 and 12 45 pH NIST mV Range 1000mV Resolution 200 200mV 0 1mV Other range 1mV Accuracy 0 2 F S Ion Range pX 0 00 10 00 Ion concentration 0 1999 Switchable unit pX mol L ppm mg L Resolution 3 4...

Page 5: ...g ATC automatic temperature compensation MTC manual temperature compensation Ion symbol Calibration guide icon Units of measurement Automatic reading hold icon Temperature units Stability icon of read...

Page 6: ...rt press In measurement mode press to turn on off the backlight In calibration mode press to calibrate In the parameter setup mode press to enter and open the sub menu confirm parameter When measureme...

Page 7: ...to change the temperature value long press for quick changing of the temperature value 3 5 2 Automatic Power off The meter is with automatic power off function Select from parameter P6 4 to select aut...

Page 8: ...rectly goes to automatic temperature compensation mode When the meter is connected to the pH electrode only it will enter manual temperature compensation mode Press or to input the temperature into th...

Page 9: ...H7 00 buffer to conduct the 1st point of calibration Dip the pH electrode and the temperature probe into pH7 00 buffer solution stir gently and let it stand still and wait for the reading to be stable...

Page 10: ...adopt 4 5 4 During the process of calibration and measurement the meter has self diagnosis functions indicating information Er1 Er4 please refer to Table 4 for relevant solutions 4 5 5 The calibratio...

Page 11: ...into 1 60pH calibration solution stir gently and let it stand still and wait for the reading to become stable When the smiley icon displays on LCD press to adjust the value to 1 60 then press to confi...

Page 12: ...ce The soaking solution contained in the supplied protective bottle is used to maintain hydration of the glass bulb and liquid junction Loose the cap take out the electrode and rinse in distilled or d...

Page 13: ...ss bulb Electrodes that have been used for a long period of time will become aged Submerge the electrode in 0 1 mol L hydrochloric acid for 24 hours then rinse the electrode in distilled water then su...

Page 14: ...quality or measuring value use ORP standard solution to test mV and see whether ORP electrode or meter works properly 5 3 2 Clean and activate ORP electrode After the electrode has been used over a lo...

Page 15: ...ing should be 1 00 x 10 3 mol L When we use the unit of pX and ppm the reading we get is the value see Diagram 7 and 9 6 1 3 Choose the ion type The meter is built in with 16 ions frequently being use...

Page 16: ...m right corner of the LCD When the reading is stable and the stable ion displays on LCD press or to adjust the value to be 1 00 E 5 1 00 10 5 press to finish the 1st point calibration the bottom left...

Page 17: ...o be rinsed and the solution contamination can be avoided 6 3 3 Calibration error reminder Ion electrode calibration may fail often The reason could be calibration solution error the quality problem o...

Page 18: ...4 Factory default setting For factory default setting please refer to parameter setting P5 3 All calibration data is deleted and the meter restores to the theory value Some functions restore to the or...

Page 19: ...n menu and P6 0 for basic parameter menu Diagram 13 Main menu of parameter setting Sub menu of pH parameter setting Sub menu of ion parameter setting Sub menu of basic parameter setting P1 1 pH buffer...

Page 20: ...ss to confirm USA USAseries nIS NIST series CUS Customized calibration 2 Press to enter P1 2 or press toreturn P1 2 Resolution 0 01 0 1 1 Press and 0 01 flashes press to select between 0 01 0 1 Press...

Page 21: ...er the ion selection interface see Diagram b 2 Press or to move the cursor press to confirm 3 Press to enter P5 2 or press to return P5 2 Select between ppm and mg L 1 Press ppm flashes press to selec...

Page 22: ...press to confirm Off Unlock up On lock up Stable reading 10 seconds and the reading will be automatically locked up 2 Press to enter P6 3 or press to return P6 3 Select backlight timing 1 2 3 Off 1 P...

Page 23: ...ttributable to responsibility of Environmental Express for a period of THREE YEARS SIX MONTHS for the probe from the delivery This limited warranty does not cover any damages due to Transportation sto...

Page 24: ...e to factory default setting No Yes No P6 0 Basic Parameter P6 1 Temperature unit P6 2 Automatic lock up timing Off On P6 3 Backlight timing 1 2 3 Off P6 4 Automatic power off timing 10 20 30 Off Appe...

Page 25: ...ort 800 745 8218 Email tech environmentalexpress com Website www environmentalexpress com Icon Self diagnosis Information pH Ion Press when measuring value is not stable during calibration Wrong pH bu...

Reviews: