MIR 9000
Environnement
S.A
5–11
JULY 2009
DISPLAY CAUSE
POSSIBLE
ACTIONS
Threshold control X
–
The measurement of the X parameter
is higher than the threshold value.
–
Compare the measurement of the X
parameter with the threshold value
programmed in the “Configuration /
Limits and Relays” menu.
Threshold control
–
The measurement is higher than the
threshold value and the analyzer is in
zero mode.
–
In the “Configuration / Limits and
Relays” menu, set the parameter with
a threshold value on the Zero relay.
–
The problem is located in the H
2
O
channel. The analyzer works with a
SEC probe.
–
Problem with drying efficiency on the
SEC probe. Check the following
points:
–
A:
Check the pressure of the
compressed air (4 bars) and the
efficiency of the ejector of the SEC
probe.
–
Check the flows in the SEC probe.
–
B: Check the Silicagel. A green
colour indicates a fault in the MDS
dryer.
–
C: change the Perma-Pure dryers.
–
The analyzer works with a dilution
module. The problem is located in the
H
2
O or CO
2
channel.
–
The threshold on H
2
O or CO
2
indicates a modification in the dilution
coefficient.
–
Check the following points:
–
A: Check the excess flow on the
flowmeter of the dilution module.
–
B: dismount and clean the dilution
module. Take good note of the
restrictors’ location .
–
Change the O-ring seals.
–
Control the rate of dilution by making
a calibration.
Summary of Contents for Envea MIR 9000
Page 10: ...Duplication prohibited MIR 9000 EnvironnementS A 0 10 JULY 2009 Page intentionally left blank...
Page 22: ...EnvironnementS A MIR 9000 Duplication prohibited 1 12 JULY 2009 Figure 1 7 Analysis cabinet...
Page 27: ...Duplication prohibited MIR 9000 EnvironnementS A 1 17 JULY 2009 Page intentionally left blank...
Page 35: ...Duplication prohibited MIR 9000 EnvironnementS A 2 5 JULY 2009 Page intentionally left blank...
Page 102: ...EnvironnementS A MIR 9000 Duplication prohibited 4 2 JULY 2009 Page intentionality left blank...
Page 157: ......
Page 159: ......
Page 161: ......
Page 163: ...EnvironnementS A MIR 9000 Duplication prohibited 6 10 JULY 2009 Figure 6 4 Oxygen probe...
Page 165: ...EnvironnementS A MIR 9000 Duplication prohibited 6 12 JULY 2009 Page intentionally left blank...
Page 180: ...EnvironnementS A MIR 9000 Duplication prohibited 6 22 JULY 2009 Page intentionally left blank...
Page 182: ...EnvironnementS A MIR 9000 Duplication prohibited 7 2 JULY 2009 Page left intentionally blank...
Page 206: ...EnvironnementS A MIR 9000 Duplication prohibited 7 28 JULY 2009 Page left intentionally blank...
Page 212: ...EnvironnementS A DAC8 board Duplication prohibited MARCH 2018 6 Page intentionally left blank...