CORRECT
Figure 5–14 – Detector signal Oscillogram MIR NH
3
/ H
2
0 version
5.3.4
ADJUSTMENT OF THE SAMPLING
Use the two measurement channels of the oscilloscope.
Connect the probe of channel 1 to Pt 17 to visualize the signals and the probe 2 to Pt 19 to visualize
the sampling. Use the inverse function of the oscilloscope for channel 2.
The sampling on the low level of the signal should be centered and adjusted as from the
Delay 1
parameter.
Sampling on the high level is adjusted as from the
Delay 2
parameter. This should be centered on the
high levels.
INCORRECT
Electronic sampling is made too far to the left of the high level.
Figure 5–15 – Sampling Oscillogram MIR NH
3
/ H
2
0 version
Summary of Contents for Envea MIR 9000
Page 10: ...Duplication prohibited MIR 9000 EnvironnementS A 0 10 JULY 2009 Page intentionally left blank...
Page 22: ...EnvironnementS A MIR 9000 Duplication prohibited 1 12 JULY 2009 Figure 1 7 Analysis cabinet...
Page 27: ...Duplication prohibited MIR 9000 EnvironnementS A 1 17 JULY 2009 Page intentionally left blank...
Page 35: ...Duplication prohibited MIR 9000 EnvironnementS A 2 5 JULY 2009 Page intentionally left blank...
Page 102: ...EnvironnementS A MIR 9000 Duplication prohibited 4 2 JULY 2009 Page intentionality left blank...
Page 157: ......
Page 159: ......
Page 161: ......
Page 163: ...EnvironnementS A MIR 9000 Duplication prohibited 6 10 JULY 2009 Figure 6 4 Oxygen probe...
Page 165: ...EnvironnementS A MIR 9000 Duplication prohibited 6 12 JULY 2009 Page intentionally left blank...
Page 180: ...EnvironnementS A MIR 9000 Duplication prohibited 6 22 JULY 2009 Page intentionally left blank...
Page 182: ...EnvironnementS A MIR 9000 Duplication prohibited 7 2 JULY 2009 Page left intentionally blank...
Page 206: ...EnvironnementS A MIR 9000 Duplication prohibited 7 28 JULY 2009 Page left intentionally blank...
Page 212: ...EnvironnementS A DAC8 board Duplication prohibited MARCH 2018 6 Page intentionally left blank...