Duplication prohibited
MIR 9000
Environnement
S.A
5-31
JULY 2009
Test points and settings of the motherboard
Address
reference
Nature of signal
PT0 Detector
signal
PT1
Peltier instruction temperature
PT2 +15VCC
test
PT3 –15VCC
test
PT4 +5VCC
test
PT5 +15VP
test
PT6
Oxygen signal or EXT.1
PT7 0V
test
PT8 Measurement
signal
PT9 Chopper
signal
PT10 Chimney
temperature
PT11
Signal before clamp
PT12 Clamp
control
PT13
GFC index test
PT14 Cubicle
temperature
PT15
Cubicle temperature (before buffer)
PT16
IR source supply voltage
PT17 AD
input
PT18 Pressure
PT19 RUN
test
PT20 STATUS
test
PT21
Flow test or EXT.3
PT22
Wheel motor power supply test
PT23
Detector signal after offset
PT24
Flow probe power supply test
PT25 Ground
Address
reference
Nature of setting
P1
Peltier instruction temperature
P2
Offset signal after amplification
P3
Gain of signal amplification line
P4
Width of clamp signal
P7
Pressure sensor gain
P8
AD gain
P9 AD
zero
P10
Detector signal offset
P11
Detector signal gain
P12 Motor
speed
P13
UV detector signal gain
P14
Flow signal gain
Summary of Contents for Envea MIR 9000
Page 10: ...Duplication prohibited MIR 9000 EnvironnementS A 0 10 JULY 2009 Page intentionally left blank...
Page 22: ...EnvironnementS A MIR 9000 Duplication prohibited 1 12 JULY 2009 Figure 1 7 Analysis cabinet...
Page 27: ...Duplication prohibited MIR 9000 EnvironnementS A 1 17 JULY 2009 Page intentionally left blank...
Page 35: ...Duplication prohibited MIR 9000 EnvironnementS A 2 5 JULY 2009 Page intentionally left blank...
Page 102: ...EnvironnementS A MIR 9000 Duplication prohibited 4 2 JULY 2009 Page intentionality left blank...
Page 157: ......
Page 159: ......
Page 161: ......
Page 163: ...EnvironnementS A MIR 9000 Duplication prohibited 6 10 JULY 2009 Figure 6 4 Oxygen probe...
Page 165: ...EnvironnementS A MIR 9000 Duplication prohibited 6 12 JULY 2009 Page intentionally left blank...
Page 180: ...EnvironnementS A MIR 9000 Duplication prohibited 6 22 JULY 2009 Page intentionally left blank...
Page 182: ...EnvironnementS A MIR 9000 Duplication prohibited 7 2 JULY 2009 Page left intentionally blank...
Page 206: ...EnvironnementS A MIR 9000 Duplication prohibited 7 28 JULY 2009 Page left intentionally blank...
Page 212: ...EnvironnementS A DAC8 board Duplication prohibited MARCH 2018 6 Page intentionally left blank...