For the analog inputs, it is necessary to specify the MUX channels used for acquisition (
MUX Channel
parameter). Channels 17 to 19 or DTP1 to DPT3 are reserved for this application. The connection is
then made direct to the interface 2 board.
The linearity parameters (
Ax2
,
Bx
and
C
) allow converting a voltage signal (0-2.5 volt) into a
measurement (0-200 °C for example).
CAUTION: The maximum voltage on these analog inputs is 2.5 volts.
The calculated values are determined from the analog inputs. Therefore, it is not necessary to specify
a MUX input or linearization.
For the pressures measured in hPa and the temperature in °C, the chimney flowrate is calculated in
KNm
3
/h (in thousandths of normal cubic metres per hour) and the speed in m/s (in metres per
second).
The units and display formats for each parameter should be configured in the “
Configuration /
Signal-Ranges-Units”
menu.
The lower part of the screen is common to each of the parameters.
The
K Pitot
coefficient corresponds to the coefficient of the Pitot tube itself in use (usual coefficient:
0.75).
The
Mass vol.
parameter corresponds to the volumetric mass of gas (usual coefficient: 1.293 g/l).
The
Chimney
parameter corresponds to the diameter of the chimney, in metres.
The
Backblow
parameter defines the frequency of back-blowings per hour (01h, 03h, 06h, 12h or
24h); the duration is defined by the
Time
parameter.
The measurements are blocked during back-blowing and a stabilisation period equal to the duration of
the back-blowing.
The back-blowing command is transferred to the EVC5 connection of the TIG2 board.
For a nil frequency (00 h), the measurements are blocked during each of the back-blowing cycles of
the MIR. This function enables back-blowing the Pitot tubes via a command of the SEC Probe.
Output
Unit:
MUX channel
Chimney flowrate
Fext – KNm
3
/h Calculated
value
-
Chimney temperature
Text - °C
Analog input
17
Chimney pressure
Pext – hPa
Analog input
18
Gas speed
Sext – m/s
Calculated value
-
Pitot pressure
Pdif – hPa
Analog input
19
Figure 3–3 – Configuration of the five DTP parameters
STOP
Summary of Contents for Envea MIR 9000
Page 10: ...Duplication prohibited MIR 9000 EnvironnementS A 0 10 JULY 2009 Page intentionally left blank...
Page 22: ...EnvironnementS A MIR 9000 Duplication prohibited 1 12 JULY 2009 Figure 1 7 Analysis cabinet...
Page 27: ...Duplication prohibited MIR 9000 EnvironnementS A 1 17 JULY 2009 Page intentionally left blank...
Page 35: ...Duplication prohibited MIR 9000 EnvironnementS A 2 5 JULY 2009 Page intentionally left blank...
Page 102: ...EnvironnementS A MIR 9000 Duplication prohibited 4 2 JULY 2009 Page intentionality left blank...
Page 157: ......
Page 159: ......
Page 161: ......
Page 163: ...EnvironnementS A MIR 9000 Duplication prohibited 6 10 JULY 2009 Figure 6 4 Oxygen probe...
Page 165: ...EnvironnementS A MIR 9000 Duplication prohibited 6 12 JULY 2009 Page intentionally left blank...
Page 180: ...EnvironnementS A MIR 9000 Duplication prohibited 6 22 JULY 2009 Page intentionally left blank...
Page 182: ...EnvironnementS A MIR 9000 Duplication prohibited 7 2 JULY 2009 Page left intentionally blank...
Page 206: ...EnvironnementS A MIR 9000 Duplication prohibited 7 28 JULY 2009 Page left intentionally blank...
Page 212: ...EnvironnementS A DAC8 board Duplication prohibited MARCH 2018 6 Page intentionally left blank...